IBM M/02-212 operating instructions A1Data sheets of standard measuring systems

Page 55

APPENDIX

A1Data sheets of standard measuring systems

Table Standard cylinder measuring systems according to DIN 53019 / ISO 3219 (consists of measuring bob and measuring cup)

For R/S Rheometer three types of measuring bobs are available:

MB-CC48 ... CC8 DIN/RC

MB-CC48 ... CC8 DIN/FTK

MBA-CC48...CC8 DIN/FTK with EMB-CC48...CC8

 

Measuring system

 

CC48

 

CC45

 

CC25

 

CC14

 

CC8

 

 

Shear rate range

 

0 ...

 

0 ...

 

0 ...

 

0 ...

 

0 ...

 

 

[s-1]

 

4,114

 

1,032

 

1,032

 

1,032

 

1,032

 

 

Shear stress range

 

0 ...

 

0 ...

 

0 ...

 

0 ...

 

0 ...

 

 

[Pa]

 

206

 

195

 

1141

 

6501

 

34,844

 

 

Viscosity range

 

0.005 ...

 

0.020 ...

 

0.118 ...

 

0.672 ...

 

3.60 ...

 

 

[Pas]

 

32

 

15

 

100

 

500

 

3,000

 

 

Filling volume

 

70

 

100

 

17

 

3

 

0.5

 

 

[ml]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Shear rate factor Kγ

 

5.142

 

1.291

 

1.291

 

1.291

 

1.291

 

 

[min/s]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Shear stress factor τ

 

0.1900

 

0.1958

 

1.1418

 

6.501

 

34.844

 

 

[Pa]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Radius of measuring bob Ri

 

23.9

 

22.5

 

12.5

 

7

 

4

 

 

[mm]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Radius of measuring cup Ra

 

24.4

 

24.4

 

13.56

 

7.59

 

4.34

 

 

[mm]

 

 

 

 

 

 

 

 

 

 

 

 

Radius of shaft Rs

 

3.5

 

3.5

 

3.5

 

2.1

 

1.2

 

 

[mm]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Angle of measuring bob

 

120

 

120

 

120

 

120

 

120

 

 

cone α

 

 

 

 

 

 

 

[°]

 

 

 

 

 

 

 

 

 

 

 

 

Distance between lower

 

 

 

 

 

 

 

 

 

 

 

 

edge of meas. bob and

 

35

 

35

 

15.5

 

13

 

12

 

 

meas. cup bottom L' [mm]

 

 

 

 

 

 

 

 

 

 

 

 

Immersion of measuring

 

22.5

 

22.5

 

12.5

 

7

 

4

 

 

shaft L'' [mm]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Length of meas. bob L

 

67.5

 

67.5

 

37.5

 

21

 

12

 

 

[mm]

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ratio of Radii δ = Error!

 

1.0209

 

1.0844

 

1.0848

 

1.0843

 

1.0850

 

 

Resistance coefficient cL

 

1.1

 

1.1

 

1.1

 

1.1

 

1.1

 

The given ranges are standard values (not maximum values).

α

M easuring geometrie DIN 53019

Brookfield Engineering Labs., Inc.

Page 55

M/02-212

Image 55
Contents Brookfield R/S Rheometer Brookfield Engineering Labs., Inc 02-212 Contents Appendix Use of the Rheometer General DescriptionMeasuring Principle II.1 R/S Rheometer II. System ConfigurationRheometer main instrument minimum configuration II.3 Computer System II.2 Measuring DevicesHints for height adjustment of the stand III. Instrument InstallationIII.1 Stand Mounting III.2 Electrical Connections III.2.1 Temperature Sensor Pt100Use data link cable of Brookfield delivery only III.2.2 AC-AdaptorIII.2.3 Printer Connection III.2.4 Computer ConnectionFTK-CC III.3 Mounting of temperature controlled measuring devicesThermostat connection to temperature control device FTK-CC III.3.2 Mounting of the measuring device ME-CP/PP Thermostat connection to temperature control device FTK-CCBrookfield Engineering Labs., Inc 02-212 Brookfield Engineering Labs., Inc 02-212 Mounting III.3.3 Mounting of the cooling device KEIV.2 Handling IV. Environment, handling, cleaning and maintenanceIV.3 Cleaning IV.1 Operating environment, StorageIV.4 Maintenance Measurement directly in the substance Measuring SystemsDepth of immersion Measurement with substance in measuring systemMeasurement with temperature control device FTK-CC Turn left Turn right Lower Lift Brookfield Engineering Labs., Inc 02-212 Rheometer Ver .02 #xxxxxx 07.05.98 Brookfield VI. Operation and Menue SystemVoltage check VCC15.25 Brookfield VI.1 KeyboardKey-function dependent to executed operation Print programs from memory VI.2 Menu System of R/S RheometerZero-point calibration Define/edit programMain-menue Run Single Run Program Remote CONFIGURATION-,emie Set time/date Set RS232 Param LanguageMenu handling VI.3 Selecting from ListsInput Values Val.1/s0200.00 Nr. of MP010 Times0100 VI.4 Input of numerical values and alphanumeric textsSelect Measuring system CC48 CC45 Input Values Val.1/s0100.00 Nr. of MP010 TimesInput Values Val.1/s0290.00 Nr. of MP Times0100 Input Values Val.1/s0200.00 Nr. of MP Times0100VI.5.1 MAIN-Menue → Run Single VI.5 Menue entries Main menueProg. Running Wait for 1.MP Output of MPs to No output device Memory Menu ST StartSelect program Prog xyz Test Prog Öl Program end Single mode Total time100s Total #MP10VI.5.2 MAIN-Menue → Run Program No valid program Enter program firstProg. running Wait for 1.MP #1 Output of MPs to Printer Memory MenueProgram end Test Total time200s Total #MP40 VI.5.5 Main menue → ConfigurationVI.5.3 Main menue → Remote VI.5.4 Main menue → UtilitiesVI.6.1 Utilities → Zero Calibration VI.6 Menue entries in the Utilities-MenueVI.6.2 Utilities → Edit Programs D1/s=ft Select Input-Mode -D1/s Nrpm TauPa Select Prog to edit Prog xyz Test NEWSelect Meas. system CC45 CC25 CC14 Nr. of stepsInput Step Nr time s Range-Error Range-Error Start-Value min 0.90 maxEdit Program name Printer not ready OK cont. ST stop VI.6.3 Utilities → Print ProgramsVI.6.4 Utilities → Measuring systems Test Press OK to saveSel. Output-device Printer RS232 VI.6.5 Utilities → Print memoryEnter Meas. Sys. #4 Name CP25 tauprom 01.1418 kgamma Press OK to save Name CC25 tauprom 01.1418 kgammaVI.6.6 Utilities → Clear memory VI.6.7 Utilities → Measure temperatureSelect output device No output device Printer RS232 VI.7 Menu entries of the Configuration menuVI.7.1 Configuration → Set output mode Service, Service2 for service personnel onlyVI.7.3 Configuration → Reset meascount VI.7.2 Configuration → MeasCount modeVI.7.4 Configuration → Set Time/Date New Time/Date 1715 Press OK to save VI.7.5 Configuration → Set RS232 ParameterRS232 4800,8,n,1 Change RS232-Par.? Okyes Stno Baudrate -4800 9600 VI.7.6 Configuration → LanguageVI.7.7 Configuration → Service, Service VI.8 Serial Data transfer via Interface RS232Brookfield Engineering Labs., Inc 02-212 Brookfield Engineering Labs., Inc 02-212 Measurement with FTK-CC VII.1 Measuring in manual modeVII. Measurements Measurement controlVII.2 Measuring in Remote-mode Measurement in remote control modeVIII. Technical data Measured or evaluated values Weight Temperature rangeIX. Guarantee Given ranges are standard values not maximum values A1Data sheets of standard measuring systemsDG DIN Table Plate/Plate measuring systems DIN Measuring system C25-1 C25-2 C50-1 C50-2 C75-1 C75-2Range error End value min 0.90 max A2 Error messagesRange Error Printer ErrorError #3 Zero cal. error Please retry cal ST stop Zero Calibration ErrorBreak torque max! Single mode Total MP Total time Break Speed MAX Break Speed max! Single mode Total MP Total timeRheometer side A3 Pin-layout of the serial data cablePC Side Other countries A4 Requirements to the mains connecting cablesGeneral information USA and CanadaEnglishGermanEnglishGerman A5 Language cross referenceGerman menue system of RC20 Index

M/02-212 specifications

The IBM M/02-212 is a cutting-edge platform that showcases the company’s dedication to excellence in technology and innovation. Designed for both enterprise performance and flexibility, this model adheres to IBM's high standards for processing power and reliability, making it a standout option for businesses looking to enhance their operations.

One of the main features of the M/02-212 is its powerful multi-core processor, which provides exceptional speed and efficiency. This architecture enables the system to handle multiple tasks simultaneously without compromising performance, making it ideal for data-intensive applications and complex processing tasks. It supports a range of workloads, from traditional enterprise applications to modern cloud-based solutions.

In terms of memory, the M/02-212 offers extensive capacity options, utilizing high-speed RAM technology that ensures swift data access and reduced latency. The combination of powerful processing and ample memory resources allows organizations to execute demanding applications and large-scale databases seamlessly.

The system is designed with advanced security features that are integral for protecting sensitive information. Encryption capabilities, secure boot options, and comprehensive identity management functionalities are built-in, providing a robust defense against potential cyber threats. This is particularly valuable for industries like finance and healthcare, where data integrity is paramount.

Moreover, the IBM M/02-212 leverages state-of-the-art virtualization technologies, enabling companies to optimize resource utilization and increase operational efficiency. Through virtualization, multiple virtual machines can run on a single physical server, allowing businesses to reduce hardware costs and improve overall system management.

The M/02-212 also emphasizes scalability, which means organizations can easily expand their capabilities as their needs evolve. This flexibility allows businesses to adapt to changing market conditions without requiring a complete overhaul of their existing infrastructure.

Additionally, it comes equipped with a user-friendly management interface, providing administrators with the tools necessary to monitor system performance, allocate resources, and troubleshoot issues efficiently.

In summary, the IBM M/02-212 combines powerful processing capabilities, robust security measures, advanced virtualization, and exceptional scalability. These characteristics make it an excellent choice for enterprises looking to enhance their IT infrastructure while ensuring efficiency and data security.