5. For the next measurement, remove the used lens filter and put on a new, clean lens filter. Clear the display by pushing the «start» button once. Fit the probe snugly into the ear canal, then push the «start» button.
The Braun ThermoScan ear thermometer turns off automatically after 60 seconds of inactivity.
Temperature taking hints
•A measurement taken in the right ear may differ from the measurement taken in the left ear. Therefore, always take the temperature in the same ear.
•The ear must be free from obstructions or excess earwax
•External factors may influence ear tempera- tures, including when an individual has:
–been lying on one ear or the other
–had their ears covered
–been exposed to very hot or very cold temperatures, or
–been recently swimming or bathing.
In these cases, remove the individual from the situation and wait 20 minutes prior to taking a temperature.
•If ear drops or other ear medications have been placed in the ear canal of one ear, take the temperature in the untreated ear.
Care and cleaning
The probe tip is the most delicate part of the thermometer. It has to be clean and intact to ensure accurate readings.
If the thermometer is accidentally used without a lens filter, clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol.
Once the alcohol has dried completely, a new lens filter can be put on and a temperature measure- ment taken. If the probe tip is damaged, contact Braun.
Use a soft, dry cloth to clean the thermometer display and product exterior. Do not use abrasive cleaners. Never submerge this thermometer in water or any other liquid.
Store the thermometer and lens filters in a dry location free from dust and contamination and away from direct sunlight.
Additional lens filters (LF 40) are available at most stores selling Braun ThermoScan.
Trouble-shooting
Situation | Solution | ||
|
|
|
|
|
| No lens filter | Attach new, clean lens |
|
| ||
|
| is attached. | filter. |
|
|
|
|
|
|
|
|
|
| Ambient | Allow the thermometer |
|
| ||
|
| temperature | to remain for 30 min- |
|
| is not within | utes in a room where |
|
| the allowed | the temperature is be- |
|
| operating | tween 10 and 40 °C |
|
| range | or 50 and 104 °F. |
|
| (10 – 40 °C or |
|
|
| 50 – 104 °F). |
|
|
|
|
|