runnIng a control tEst:

There are times when you will want to do a quality control test so you know that your system is working properly. You may also want to do a control test to check the way you do your test. Simply follow these steps:

Push Out Test Strip

1.Hold the meter comfortably in one hand. Firmly grasp the meter handle with the other hand (shown here). Now, just “Pull and Push”—“Pull” the handle out until it stops and “Push” it back in until it stops again.

NOTE: After the meter handle is pushed in, the release button is UP, a test strip is pushed out and a series of “apply blood” prompts appears. The meter is now ready to run a control test.

2.Squeeze a small drop of

Breeze control solution onto a nonabsorbent surface (such as a clean piece of wax paper). Do not apply control solution to the test strip directly from the bottle.

NOTE: The control solution contains a red dye and may stain.

0

Bayer Diabetes Care Helpline: 1-800-268-7200

Image 24
Contents User Guide As Easy as Table of Contents Technical Information Introduction Look at Your Breeze Meter Down Buttons Explanation of Display Symbols Here is what each display element means TEstIng matErIals Open Meter Open Test Disc PackageInsert Disc Close MeterCounting the Number of Strips Left RunnIng a Blood glucosE tEst Prepare Lancing DevicePush Out Test Strip Stick your finger and form a small round drop of blood Test  Bayer Diabetes Care Helpline Strip may not release properly Precautions Marking a Deleted Blood Test Symptoms of High or Low Blood Glucose Expected Blood Glucose ValuesUnusual Test Results RunnIng a control tEst Test Marking a Control Test Page BEyond EvEryday usE REcallIng tEst rEsultsViewing Your Averages Clearing Stored Test Results SEtuP functIon Setting the TimeSetting the Date Page Setting the Buzzer Level Setting the Time Format 12- or 24-hour Clock Setting the Date Format TransfErrIng rEsults to a comPutEr ClEanIng your mEtEr REPlacIng thE BattEry To replace the battery, follow these steps Page Technical Information Solving Problems Troubleshooting and Error Codes What You See What It Means What You Should Do What You See What It Means What You Should Do This is a normal expected action Control solution with an acceptable Customer Service Replacement Items Lifetime Warranty Bayer Diabetes Care Helpline Manufactured for