What You See |
| What It Means | What You Should Do |
|
|
|
|
Control Test | • | Control solution is | Run another control test. If the |
result is out of |
| past its expiration | result is still out of range, retest |
range (too high |
| date or is past the | with a new |
or too low). |
| control solution with an acceptable | |
| expiration date and open | ||
|
| ||
| • | date. If still out of range, contact | |
| the Bayer Diabetes Care Helpline. | ||
|
| expiration date. | |
|
|
| |
| • | Deteriorated test |
|
|
| strip due to heat or |
|
|
| exposure to moisture. |
|
| • | Control solution |
|
|
| is not at room |
|
|
| temperature. |
|
| • Not enough control |
| |
|
| solution drawn into |
|
|
| test strip. |
|
|
|
| |
Meter begins | The inside of your | Wick the mosture away (see | |
countdown before | meter may be wet. | page 34). If you continue to | |
blood is applied. |
|
| experience difficulty, contact |
|
|
| the Bayer Diabetes Care Helpline. |
|
|
| |
Some display | This may affect the way | If there is a difference in your | |
segments do not | you see your results. | meter’s display and the one | |
appear to be |
|
| shown on page 5, contact the |
showing when |
|
| Bayer Diabetes Care Helpline. |
is held in or |
|
|
|
the meter handle |
|
|
|
is all the way out. |
|
|
|
|
|
|
|
44 | Bayer Diabetes Care Helpline: |