Bayer HealthCare 2 manual What You See What It Means What You Should Do

Page 45

What You See

What It Means

What You Should Do

 

 

 

 

 

 

 

 

Low Battery symbol

Replace the battery (CR2032)

 

 

 

flashes if battery power

within 20 readings or one week.

 

 

 

is low.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Battery life has ended.

Replace the battery (CR2032).

Temperature is outside

Allow the meter to reach room

the operating range of

temperature.

the meter (too hot or

 

too cold).

 

 

 

Meter does not

Remove the test disc and inspect

recognize the test disc.

the meter for foreign material.

 

Rotate the disc and reinstall it. If

 

error repeats, try inserting a new

 

disc. If error continues, contact the

 

Bayer Diabetes Care Helpline.

 

 

Test strip was removed

Take care to not press the release

during the test, or the

button or remove the test strip

release button was

during the test. Release the used

pressed during the test.

test strip and repeat the test

 

(see page 10). If the error contin-

 

ues, contact the Bayer Diabetes

 

Care Helpline.

The meter was open

Close the meter and push the

when the meter handle

handle in.

was pulled.

 

(with continuous beep)

www.ascensia.ca41

Image 45
Contents User Guide As Easy as Table of Contents Technical Information Introduction Look at Your Breeze Meter Down Buttons Explanation of Display Symbols Here is what each display element means TEstIng matErIals Open Test Disc Package Open MeterClose Meter Insert DiscCounting the Number of Strips Left Prepare Lancing Device RunnIng a Blood glucosE tEstPush Out Test Strip Stick your finger and form a small round drop of blood Test  Bayer Diabetes Care Helpline Strip may not release properly Precautions Marking a Deleted Blood Test Expected Blood Glucose Values Symptoms of High or Low Blood GlucoseUnusual Test Results RunnIng a control tEst Test Marking a Control Test Page REcallIng tEst rEsults BEyond EvEryday usEViewing Your Averages Clearing Stored Test Results Setting the Time SEtuP functIonSetting the Date Page Setting the Buzzer Level Setting the Time Format 12- or 24-hour Clock Setting the Date Format TransfErrIng rEsults to a comPutEr ClEanIng your mEtEr REPlacIng thE BattEry To replace the battery, follow these steps Page Technical Information Solving Problems Troubleshooting and Error Codes What You See What It Means What You Should Do What You See What It Means What You Should Do This is a normal expected action Control solution with an acceptable Customer Service Replacement Items Lifetime Warranty Bayer Diabetes Care Helpline Manufactured for