Bayer HealthCare 2 manual Troubleshooting and Error Codes

Page 44

Troubleshooting and Error Codes:

When you have a problem or a question concerning what you see on your screen, the following may be helpful.

What You See

What It Means

What You Should Do

 

 

 

 

 

 

 

 

Test result is below

Repeat the test and make sure

 

 

 

0.6 mmol/L.

the test strip is completely filled.

 

 

 

 

If LO appears again, contact

 

 

 

 

 

 

 

 

 

 

 

 

your physician or healthcare

 

 

 

 

professional.

 

 

 

 

 

 

 

 

Test result is above

Wash your hands and the

 

 

 

33.3 mmol/L.

test site. Repeat the test. If

 

 

 

 

HI appears again, contact

 

 

 

 

 

 

 

 

 

 

 

 

your physician or healthcare

 

 

 

 

professional.

 

 

 

 

 

 

 

 

You have not pushed

Push the meter handle in until it

 

 

 

the meter handle all

clicks. You must push the handle

 

 

 

the way in.

all the way in and see the “apply

 

 

 

 

 

 

 

 

 

 

blood” prompts before you touch

 

 

 

 

the strip to the blood drop.

 

 

 

 

 

 

 

 

• The 10-Test Disc is

Insert a new 10-Test Disc. If a full disc

 

 

 

out of test strips.

is already inserted, remove and in-

 

 

 

• There is no disc in

spect for foreign material. Rotate the

 

 

 

the meter.

disc and reinstall it, bumpy side up.

 

 

 

• The disc has been

If error repeats, try inserting a new

 

 

 

inserted upside down.

disc. If the error continues, contact

 

 

 

• The disc is defective.

the Bayer Diabetes Care Helpline.

 

 

 

 

 

 

 

 

The meter will mark

Test only when the meter tempera-

 

 

 

a test result with a

ture is between 10°C and 45°C.

 

 

 

thermometer and an

 

 

 

 

 

 

 

 

 

 

 

 

if the meter is too cold

 

 

 

 

or too hot when the test

 

 

 

 

is performed. The result

 

 

 

 

will not be included in

 

 

 

 

any averages calculated.

 

 

 

 

 

 

40

Bayer Diabetes Care Helpline: 1-800-268-7200

Image 44
Contents User Guide As Easy as Table of Contents Technical Information Introduction Look at Your Breeze Meter Down Buttons Explanation of Display Symbols Here is what each display element means TEstIng matErIals Open Meter Open Test Disc PackageInsert Disc Close MeterCounting the Number of Strips Left RunnIng a Blood glucosE tEst Prepare Lancing DevicePush Out Test Strip Stick your finger and form a small round drop of blood Test  Bayer Diabetes Care Helpline Strip may not release properly Precautions Marking a Deleted Blood Test Symptoms of High or Low Blood Glucose Expected Blood Glucose ValuesUnusual Test Results RunnIng a control tEst Test Marking a Control Test Page BEyond EvEryday usE REcallIng tEst rEsultsViewing Your Averages Clearing Stored Test Results SEtuP functIon Setting the TimeSetting the Date Page Setting the Buzzer Level Setting the Time Format 12- or 24-hour Clock Setting the Date Format TransfErrIng rEsults to a comPutEr ClEanIng your mEtEr REPlacIng thE BattEry To replace the battery, follow these steps Page Technical Information Solving Problems Troubleshooting and Error Codes What You See What It Means What You Should Do What You See What It Means What You Should Do This is a normal expected action Control solution with an acceptable Customer Service Replacement Items Lifetime Warranty Bayer Diabetes Care Helpline Manufactured for