53IIB/54IIB

Users Manual

3.Press to step through the minimum, maximum, average, and current logged reading. For example, Figure 4 shows the maximum reading in memory. The maximum reading occurred at 1:49 P.M. and was stored in memory location 5.

Communicating with a PC

You can transfer the contents of the thermometer’s memory to a PC using FlukeView Forms. The communication requires an IR (infrared) serial connection. Refer to the FlukeView Forms Installation Guide and FlukeView Help.

FlukeView Forms places the logged readings into standard (default) or customized forms. The forms also display user comments. You can use these forms to satisfy ISO-9000 documentation requirements.

When you send logged readings to a PC, all functions are disabled except power off, backlight, and turning off communication.

Figure 4. Maximum Reading

4.Press to stop viewing logged readings.

Note

The thermometer calculates the minimum and maximum of all logging sessions in memory.

The display shows "- - : - -" if the log is empty.

aat04f.eps

Press , (PC/IR SEND) to toggle the IR port on and off.

The display shows:





When the IR port is enabled you can communicate with FlukeView Forms.

18

Image 22
Contents 53IIB/54IIB Limited Warranty and Limitation of Liability Table of Contents 53IIB/54IIB Safety Information Contacting Fluke53IIB/54IIB Safety Information International Symbols Getting StartedComponents ComponentsDisplay Elements Display ElementsButtons ButtonsPress to freeze or unfreeze the displayed readings Changing Setup Options Using the ThermometerEntering and Exiting Setup Changing the Logging Interval Changing the Thermocouple Type Changing the OffsetEnabling or Disabling Sleep Mode Setting the TimeConnecting a Thermocouple Measuring TemperaturesChanging the Line Frequency Holding the Displayed Temperatures Displaying TemperaturesViewing the MIN, MAX, and AVG Readings Using Memory Using the Offset to Adjust for Probe ErrorsInitial Conditions and Data Entries Starting and Stopping LoggingClearing Memory Viewing Logged ReadingsCommunicating with a PC Environmental Cleaning the Case and Holster MaintenanceSpecifications Replacing the Batteries80 PK-1 Thermocouple supplied with thermometer ElectricalReplacement Parts and Accessories Model53IIB/54IIB