Fluke 54IIB, 53IIB specifications Changing the Thermocouple Type, Changing the Offset

Page 15

53IIB/54IIB

Changing Setup Options

Changing the Thermocouple Type

1.Press or until the display shows TYPE.

2.Press to display the thermocouple type choices. The currently selected thermocouple blinks.

3.Press or until the thermocouple you want appears on the display.

4.Press to store the thermocouple type in memory.

Changing the Offset

You can adjust the thermometer's readings to compensate for the errors of a specific thermocouple. See "Using the Offset to Adjust for Probe Errors." The allowable adjustment range is ±5.0 oC or K, and ±9.0 oF.

Model 54: You can store individual offsets for T1 and T2.

1.Press or until the display shows and T1 or T2.

2.Press to indicate that you want to change the offset setting.

The temperature measurement plus the offset appears in the primary display. The offset appears in the secondary display.

3.Press or until the primary display shows the correct reading.

4.Press to store the offset setting in memory.

Remember to reset the offset to 0.0 when it is no longer needed. The offset automatically resets to 0.0 when you change the thermocouple type.

11

Image 15
Contents 53IIB/54IIB Limited Warranty and Limitation of Liability Table of Contents 53IIB/54IIB Contacting Fluke Safety Information53IIB/54IIB Safety Information Getting Started International SymbolsComponents ComponentsDisplay Elements Display ElementsButtons ButtonsPress to freeze or unfreeze the displayed readings Using the Thermometer Changing Setup OptionsEntering and Exiting Setup Changing the Logging Interval Changing the Offset Changing the Thermocouple TypeSetting the Time Enabling or Disabling Sleep ModeMeasuring Temperatures Connecting a ThermocoupleChanging the Line Frequency Displaying Temperatures Holding the Displayed TemperaturesViewing the MIN, MAX, and AVG Readings Using the Offset to Adjust for Probe Errors Using MemoryStarting and Stopping Logging Initial Conditions and Data EntriesViewing Logged Readings Clearing Memory Communicating with a PCReplacing the Batteries MaintenanceSpecifications Environmental Cleaning the Case and HolsterElectrical 80 PK-1 Thermocouple supplied with thermometerModel Replacement Parts and Accessories53IIB/54IIB