Fluke 53IIB, 54IIB Initial Conditions and Data Entries, Starting and Stopping Logging

Page 20

53IIB/54IIB

Users Manual

Initial Conditions and Data Entries

Logged readings include initial conditions and data entries.

The initial conditions are the thermocouple type and the offsets for each thermocouple input. You can only view initial conditions using FlukeView Forms.

The data entries are a time stamp, the T1 reading, and the T2 and T1-T2 readings (Model 54). You can view these values by pressing or using FlukeView Forms. Temperature readings display 0.1 degree resolution in FlukeView Forms.

The thermometer has 500 memory locations. The thermometer stores 499 sets of temperature readings and one set of initial conditions when logging continuously. It stores 250 sets of temperature readings and 250 sets of initial conditions when logging individual points manually.

Starting and Stopping Logging

Setup, memory clear, and PC communications are inaccessible during logging.

1.Set the logging interval. (See “Changing Setup Options.”)

2.Press to start logging. The display shows .

3.Press again to stop logging.

4.If you selected a manual logging interval, press each time you want to store logged readings in memory.

16

Image 20
Contents 53IIB/54IIB Limited Warranty and Limitation of Liability Table of Contents 53IIB/54IIB Safety Information Contacting Fluke53IIB/54IIB Safety Information International Symbols Getting StartedComponents ComponentsDisplay Elements Display ElementsButtons ButtonsPress to freeze or unfreeze the displayed readings Entering and Exiting Setup Using the ThermometerChanging Setup Options Changing the Logging Interval Changing the Thermocouple Type Changing the OffsetEnabling or Disabling Sleep Mode Setting the TimeChanging the Line Frequency Measuring TemperaturesConnecting a Thermocouple Viewing the MIN, MAX, and AVG Readings Displaying TemperaturesHolding the Displayed Temperatures Using Memory Using the Offset to Adjust for Probe ErrorsInitial Conditions and Data Entries Starting and Stopping LoggingClearing Memory Viewing Logged ReadingsCommunicating with a PC Maintenance SpecificationsEnvironmental Cleaning the Case and Holster Replacing the Batteries80 PK-1 Thermocouple supplied with thermometer ElectricalReplacement Parts and Accessories Model53IIB/54IIB