Programming Guide
Description | Char | Format | P | B |
| S | Legal values | Factory | LCD |
|
|
|
|
|
|
|
| default |
|
Presel. setpoint / relay | SV | nnn.n |
|
|
|
| (1) |
|
|
function |
|
|
| * |
|
|
|
|
|
Target temperature | T | nnn.n | * |
|
| in current scale |
|
| |
|
|
|
| * |
|
| (°C / °F) |
|
|
Temperature unit | U | X | * |
| * | C / F | C | U | |
Poll / Burst mode | V | X | * |
|
| * | P = poll | Poll mode |
|
|
|
|
|
|
|
| B = burst |
|
|
Burst string contents | X$ |
| * |
|
|
|
|
|
|
Multidrop address | XA | nnn | * |
|
| * | 000 – 032 | unchanged |
|
|
|
|
|
|
|
| 000 = single unit | (preset: 0) |
|
Bottom temperature of | XB | nnn.n | * |
|
|
|
|
|
|
range |
|
|
|
|
|
|
|
|
|
Restore factory defaults | XF |
|
| * |
| * |
|
|
|
Transmission | XG | n.nnn | * |
| * | 0.100 - 1.000 | 1.000 | T | |
High temperature of | XH | nnn.n | * |
|
|
|
|
|
|
range |
|
|
| * |
|
|
|
|
|
Sensor initialization | XI | n | * |
| * | 1 after Reset, |
|
| |
|
|
|
| * |
|
| 0 if XI = 0 |
|
|
Box temperature | XJ | nnn.n | * |
|
| in current |
|
| |
|
|
|
|
|
|
| scale(°C / °F) |
|
|
FTC 3 trigger / hold | XN | X | * |
|
| * | T = trigger | T |
|
|
|
|
|
|
|
| H = hold |
|
|
Analog output mode | XO | n | * |
|
| * | 0 = 0 – 20 mA | 9 |
|
|
|
|
|
|
|
| 4 = 4 – 20 mA |
|
|
|
|
|
|
|
|
| 5 = TCJ |
|
|
|
|
|
|
|
|
| 6 = TCK |
|
|
|
|
|
|
|
|
| 9 = mV |
|
|
Firmware revision | XR |
| * |
|
|
| e.g. 1.01 |
|
|
Setpoint / relay function | XS | nnn.n | * |
|
| * | (1) | 250°C |
|
(10) |
|
|
| * |
|
|
|
|
|
Trigger | XT | n | * |
|
| 0 = inactive | 0 |
| |
|
|
|
|
|
|
| 1 = active |
|
|
Unit identification | XU |
| * |
|
|
| e.g. !XUMILT4 |
|
|
Serial number | XV |
| * |
|
|
| e.g. 98123 |
|
|
Advance hold hysterese | XY | nnn.n | * |
|
| * |
|
|
|
(4) |
|
|
|
|
|
|
|
|
|
Head calibration (9) | XZ |
| * |
|
| * |
|
|
|
(1)LT: ‐40 to 600°C (‐40 to 1112°F)
(2)n = number, X = uppercase letter
MI | 79 |