Page 20

“i” Returns general information regarding the instrument, as well as some parameters that have been programmed previously in the thermometer (either by RS-232 commands or via the keypad):

Ex:

Model:

 

HHTFO101

 

Option:

 

NB Channel: 1

 

Mode:

 

Continuous Scanning Mode

 

Scanning rate:

 

Fast

 

 

Serial:

 

HHTFO111A

 

Internal Software:

V1.21

 

 

Last Factory Calibration*: 05/02/04 (YY/MM/DD)

 

Unit

 

 

oC

 

 

Keyboard Status:

Unlock

 

Channel

Zero

Span

Enabled

Temperature offset*

1

 

xxx.x

xxx.x

Yes/No

±xxxx.x

*: Temperature differential that is a consequence of the “f” command.

“q” “q” allows for selecting the fast (q+) or the slow (q-) acquisition mode. This is equivalent to the Sampling Rate selection that is available on the FOB100 keypad. In slow mode, the last 4 temperatures are averaged before being displayed or sent to the RS-232 port; this provides a better reading at the expense of being slower.

“t” or “t[i]” Returns the current temperature reading for channel # i. “---.-” is displayed if the specified channel number is not currently enabled or when no valid temperature is available for the requested channel. However, during the initial warm-up time, the error “Err1” is sent in response to the “t” command. For multi- channel instruments, if the “t” command is received without parameter, the temperatures for all channels are returned. All temperatures are returned in oC or oF unit, as specified by the “u” command.

Ex: “t4 E“ will return the temperature for channel # 4. “t E”, with the HHTFO101 thermometer, will return:

+24.3<CR>

*

The “t” command is also available in a special “ta” version, that can be useful for continuously and automatically monitoring all enabled channels. This special mode is recommended when automatically acquiring temperature readings with a suitable data-logging software, and can be activated by issuing the “ta+” command, or deactivated with the “ta-“ command. Once activated, the following information is automatically sent when the temperature on a channel has been acquired:

“C:I;T: ±xxx.x<CR>”

where i is the channel number, and xxx.x is the temperature reading (signed number).

“u[i]” Set units of measurements to either Celsius (i=c) or Fahrenheit (i=f). Note that this unit selection is the same as the temperature unit selection made from the keypad (if the display option is installed).

Ex:

“uc E“, to select ºC (Celsius).

“y” The “y” command can be used to confirm the “signal strength” of a probe and/or its connection. It can be used as a diagnostic tool to confirm the good operation of a probe and its patch cord (if applicable) connected to a channel. The returned value can be between 0.3 to approximately 3.00; a higher value is better. A minimum of 1.25 is required for a channel to return a temperature.

Ex: “y E”, with the HHTFO101 thermometer, will return:

CH1: power:2.61, lamp attenuation: 250, CCD time: 50ms (fixed)<CR>

*: 0.3 to 1.1 indicates that no probe is detected for this channel.

20

Image 20
Contents Hhtfo Page Introduction Page Introduction Hhtfo product specifications VDCCalibration Unpacking OptionsInstalling batteries Quick IntroductionMaking your first meaasurements Keypad operation Hhtfo Thermometer Hardware ReferennceDisplay description Logging menu Using the backlightSetup menu File naming convention Format sub-menuEnable sub-menu Starting and stopping a logging sessionIcon definition Rate sub-menu10 RS232 communication interface Power jackProbe connector Using Temperature Probes DescriptionPage RS-232 help menu Using Hyper Terminal in WindowsSerial Communication Description Detailed description of RS-232 commands HHTFO101HHTFO111A Typical temperature reading sequence Error Codes Analog Out DescriptionInstalling FOB100-SOFT Getting started with FOB100-SOFTQuick tour of FOB100-SOFT Hardware and software requirementsWorking with the Configuration Window Adjusting a temperature offset Acquiring and viewing temperature data Logging temperature data to a fileYour mouse to widen this column Viewing temperature data Displaying temperature trendsDisplaying analog values dials Displaying analog thermometersDisplaying digital values FOB100 Console FOB100-SOFT functions Console help menuFile menu Find Instruments RefreshWindow menu Help Menu