Page 17

Section 3

Calibration

 

 

3.1 Calibration

Units with serial numbers beginning with the letter N use the calibration procedure outlined below. Older units do not display version numbers and use the calibration procedures on page 19 .

The calibration procedure should be performed by a qualified instrumentation service technician. The indicator may be brought to any instrumentation calibration and repair laboratory (metrology lab) for the service (this manual should accompany the instrument). This procedure should be done at least once a year to guarantee continued performance within the specification.

The instrument should be calibrated at an ambient temperature of 22° C with a thermocouple simulator

14

Image 17
Contents Series Servicing North America Servicing Europe Calibration DescriptionOperation General Description Section DescriptionUnpacking Physical DescriptionCmrr Specifications±.45F±0.1% rdg Range Table Principles of Operation Section OperationReplacing the battery Recesses or each corner of the case bottom Page Page Operation Features Maintenance Calibration Section CalibrationPage For T/C type Verify calibration once finished Ra Full Scale Adj Rb Zero Adj Rc +Full Scale Adj Calibration- Thermocouple Type K, J Page Page Calibration- Thermistor Type Page Calibration- Platinum RTD Page Calibration is complete Zero CalibrationWARRANTY/DISCLAIMER Temperature Data Acquisition