at the “-1.23V” testpoint: disregarding polarity indication, record the voltage.

f.Move the DVM HI lead to the “3.7V” testpoint and record the voltage (disregard polarity).

g.The voltage measured at the “3.7V” test- point should be 3.007 times larger than the voltage measured at the “-1.23V” test- point. Adjust Rc (which varies the +3.7V line to attain the proper ratio.

h.Adjust Rb for an indicator display of 0.2°C.

i.Set the resistance box to 148.4 ohms.

j.Adjust Ra for an indicator display of +101.0°C.

k.Reassemble the indicator.

l.Perform the zero calibration

23

Image 26
Contents Series Servicing North America Servicing Europe Calibration DescriptionOperation Section Description General DescriptionPhysical Description UnpackingSpecifications Cmrr±.45F±0.1% rdg Range Table Section Operation Principles of OperationReplacing the battery Recesses or each corner of the case bottom Page Page Operation Features Maintenance Section Calibration CalibrationPage For T/C type Verify calibration once finished Ra Full Scale Adj Rb Zero Adj Rc +Full Scale Adj Calibration- Thermocouple Type K, J Page Page Calibration- Thermistor Type Page Calibration- Platinum RTD Page Zero Calibration Calibration is completeWARRANTY/DISCLAIMER Temperature Data Acquisition