c.Remove the bottom case of the indicator.

d.Adjust the voltage calibrator for an output of 0.000mV and connect the minus lead from the calibrator to “GND” and the positive lead from the calibrator to “;+IN”, on the indicator printed circuit board.

e.Disable the reference junction by jumpering together the “REF J” and “GND” testpoints, using a jumper lead terminated in mini-clips.

f.Adjust Rb for an indicator display of 0.0°C.

g.Connect the DVM LO lead to “GND”, on the indicator printed circuit board.

h.Using HI lead of DVM, measure voltage at the “-1.23V” testpoint: disregarding polarity indication, record the voltage.

i.Move the DVM HI lead to the “3.7V” testpoint and record the voltage (disregard

20

Image 23
Contents Series Servicing North America Servicing Europe Calibration DescriptionOperation General Description Section DescriptionUnpacking Physical DescriptionCmrr Specifications±.45F±0.1% rdg Range Table Principles of Operation Section OperationReplacing the battery Recesses or each corner of the case bottom Page Page Operation Features Maintenance Calibration Section CalibrationPage For T/C type Verify calibration once finished Ra Full Scale Adj Rb Zero Adj Rc +Full Scale Adj Calibration- Thermocouple Type K, J Page Page Calibration- Thermistor Type Page Calibration- Platinum RTD Page Calibration is complete Zero CalibrationWARRANTY/DISCLAIMER Temperature Data Acquisition