Calibration- Thermocouple Type K, J and

T

and E

Equipment List

4 ½ digit voltmeter

Voltage Calibrator:

Resolution, 1uV; accuracy ±0.01% (±2uV)

Ice Bath, 0.00 °C ± 0.05°C

Jumper lead terminated in mini-clips

For optimum results, the calibration should be performed with an ambient temperature from 20° 24°C (68° to 75°F). Refer to the figure on the previous page for the locations of all testpoints and adjustments.

a.Turn the indicator ON by pressing the °C key.

b.For types K, J, E and T units only, switch the indicator to the 0.1° range.

19

Image 22
Contents Series Servicing North America Servicing Europe Operation DescriptionCalibration Section Description General DescriptionPhysical Description UnpackingSpecifications Cmrr±.45F±0.1% rdg Range Table Section Operation Principles of OperationReplacing the battery Recesses or each corner of the case bottom Page Page Operation Features Maintenance Section Calibration CalibrationPage For T/C type Verify calibration once finished Ra Full Scale Adj Rb Zero Adj Rc +Full Scale Adj Calibration- Thermocouple Type K, J Page Page Calibration- Thermistor Type Page Calibration- Platinum RTD Page Zero Calibration Calibration is completeWARRANTY/DISCLAIMER Temperature Data Acquisition