polrity).

j.The voltage measured at the “3.7V” test- point should be 3.007 times larger than the voltage measured at the “-1.23V” test- point. Adjust Rc (which varies the +3.7V line to attain the proper ratio.

k.Switch the indicator to the 1° range.

l.Adjust the voltage calibrator for an output of

+53.782mV for type K; +57.942mV for type J; +75.983mV for type E; +18.824mV for type T;

m.Adjust Ra for an indicator display of +995°C for type E, or 1000°C for type J, or +1340°C for type K, or +367°C for type T.

n.Reassemble the indicator.

o.Perform the zero calibration.

21

Image 24
Contents Series Servicing North America Servicing Europe Description OperationCalibration Section Description General DescriptionPhysical Description UnpackingSpecifications Cmrr±.45F±0.1% rdg Range Table Section Operation Principles of OperationReplacing the battery Recesses or each corner of the case bottom Page Page Operation Features Maintenance Section Calibration CalibrationPage For T/C type Verify calibration once finished Ra Full Scale Adj Rb Zero Adj Rc +Full Scale Adj Calibration- Thermocouple Type K, J Page Page Calibration- Thermistor Type Page Calibration- Platinum RTD Page Zero Calibration Calibration is completeWARRANTY/DISCLAIMER Temperature Data Acquisition