with a minimum accuracy of ±0.1 degree Fahrenheit or ±0.2 degree Centigrade.

Step 1. Ensure the unit is turned off. Connect the output of the thermocouple simulator to the input of the indicator using the appropriate thermocouple wire. Set the simulator output to 0°C

Step 2. Locate the calibration switch on the back of the case labeled “Zero”. Remove the black button. Using a small screw drive or flat end of the trimmer pot, press the cal switch once. The display should read version and then switch to ambient temperature; press the switch one more time. The display should indicate .0.0 °C

Step 3. Press the switch one more time, the display should read max calibration temperature (for K type: 13.2.0, for J type: 9.6.0, for T type: 3.5.0, and for E type: 9.5.0) Now, set the simulator output to display temperature as show in table below.

15

Image 18
Contents Series Servicing North America Servicing Europe Description OperationCalibration Section Description General DescriptionPhysical Description UnpackingSpecifications Cmrr±.45F±0.1% rdg Range Table Section Operation Principles of OperationReplacing the battery Recesses or each corner of the case bottom Page Page Operation Features Maintenance Section Calibration CalibrationPage For T/C type Verify calibration once finished Ra Full Scale Adj Rb Zero Adj Rc +Full Scale Adj Calibration- Thermocouple Type K, J Page Page Calibration- Thermistor Type Page Calibration- Platinum RTD Page Zero Calibration Calibration is completeWARRANTY/DISCLAIMER Temperature Data Acquisition