Omega Vehicle Security 450 manual ±.45F±0.1% rdg

Page 8

Maximum Common Mode Voltage: 1000VAC(RMS), ±2000V Peak

Accuracy at 22°C (72°F)

 

1° range:

±0.8°C±0.1% rdg

 

±1.5°F±0.1% rdg

0.1° range:

±0.25°C±0.1% rdg

 

±.45°F±0.1% rdg

Stability with time:

 

90 days:

add ±0.05% rdg.

1 year:

add ± 0.1% rdg.

Stability with temperature: Thermocouple:

zero: ±0.03°C/°C

span: ±0.02% rdg/°C (max)

reference junction tracking ±1.0°C (±1.8°F) from 5°C to 45°C (41°F to 113°F)

Platinum RTD: ±0.02°C, ±0.01% rdg/°C

Thermistor: positive- ±0.02°C/°C negative-±0.04°C/°C

5

Image 8
Contents Series Servicing North America Servicing Europe Calibration DescriptionOperation Section Description General DescriptionPhysical Description UnpackingSpecifications Cmrr±.45F±0.1% rdg Range Table Section Operation Principles of OperationReplacing the battery Recesses or each corner of the case bottom Page Page Operation Features Maintenance Section Calibration CalibrationPage For T/C type Verify calibration once finished Ra Full Scale Adj Rb Zero Adj Rc +Full Scale Adj Calibration- Thermocouple Type K, J Page Page Calibration- Thermistor Type Page Calibration- Platinum RTD Page Zero Calibration Calibration is completeWARRANTY/DISCLAIMER Temperature Data Acquisition