Product description
1Protective cap
2Lens filter
3Probe
4«start» button
5Lens filter detector
6Display
7Battery door
How to use your Braun ThermoScan
ThermoScan
20LF
1.To achieve accurate measurements, make sure a new, clean lens filter (2) is in place before each measure- ment.
2.To turn the thermometer on, push the «start» button (4).
During an internal self- check, the display (6) shows all segments. Then the last temperature taken will be displayed together with «MEM».
As soon as this ready symbol appears, you can start the measuring process.
3.Perform an ear tug to straighten the ear canal. This gives the thermometer a clear view of the eardrum.
An ear tug is best performed by using your free hand to grasp the outer edge of the top half of the ear. To take your own temperature, wrap your free hand around the back of your head and tug your ear from behind.
Children under 1 year:
Pull the ear straight back.
Children aged 1 year to adult:
Pull the ear up and back.
4. While tugging the ear, fit the probe (3) snugly into the ear canal, then push the «start» button. A short beep will signal the start of the measuring process.
After one second a longer beep will signal the end of the measuring process. The result is shown on the display (6).
4