5. TO CLEAN THE BATH OR TO ADD WATER

SAFETY : NEVER ADD WATER OR MOVE

THE BATH IF THE POWER SWITCH (ON

THE LEFT SIDE PANEL) IS TURNED ON.

RECOMMENDATION : FOR MAXIMUM SAFETY, UNPLUG THE UNITS POWER CORD FROM THE ELECTRICAL OUTLET.

5.1TURN THE POWER SWITCH (ON THE LEFT SIDE PANEL) TO OFF POSITION TO TURN THE UNIT POWER OFF.

BEFORE REMOVING THE BATH FILLED WITH WATER, UNPLUG THE UNIT FROM THE ELECTRICAL OUTLET.

5.2Depending on usage, the water should be changed daily but not less than once per week.

5.3Remove the Plasma Bag Holder.

5.4Tilt the unit to the up position by lifting the handle in front of the control panel. With unit in up position, tighten the clamp knob located on the top left-hand side of the device (turn the

knob clockwise.) DO NOT OVER TIGHTEN. The unit will now be held in the up position.

It will be helpful at this point to have a cloth towel ready to cleanup dripping water. Or, by waiting a few minutes, the pump and heater area can drain into the water bath.

5.5GENESIS 2002: Carefully remove the water bath from the device. The bath weighs approximately twenty (20) pounds

10

Image 10
Contents Genesis 2002TM Genesis 2006TM Genesis 2006 Shown GenesisBPS, LLC Contents Unpacking To Thaw Blood ProductsPage Page Page To Remove a Thawed BAG To Maintain the Device in the READY-TO- Thaw StatusPage Page Control Panel and Indicators Genesis 2002TM Control Panel Genesis 2006TM Control Panel Control Keys Indicators Unit does not rock Pump does not circulate the water Special Features Unique SafeguardsSpecifications ORDERING/RETURN Information Ordering Information Return InformationPage Warranty Plasma Thawing Water Bath System