3.TO REMOVE A THAWED BAG

3.1Press the ROCK key to stop the rocking motion.

a. Remove the thawed bag (and any weights used) from the Plasma Bag Holder. Discard plasma thawing overwrap bag after each use, as they are designed for single use only.

b. If there is another bag being thawed, press the ROCK key to resume the rocking motion for the remaining bag.

4.TO MAINTAIN THE DEVICE IN THE READY-TO- THAW STATUS

4.1Always keep the HEAT indicator illuminated.

4.2Always keep the SET TEMP to the desired setting. (37 °C or other desired temperature). If a desired temperature of 50 °C or greater is needed, limit the time the unit is in the ready to use state. At higher temperatures, greater evaporation occurs.

4.3Always keep the ROCK indicator off (not illuminated) when not thawing plasma or other products.

9

Image 9
Contents Genesis 2002TM Genesis 2006TM Genesis 2006 Shown GenesisBPS, LLC Contents To Thaw Blood Products UnpackingPage Page Page To Maintain the Device in the READY-TO- Thaw Status To Remove a Thawed BAGPage Page Control Panel and Indicators Genesis 2002TM Control Panel Genesis 2006TM Control Panel Control Keys Indicators Unit does not rock Pump does not circulate the water Unique Safeguards Special FeaturesSpecifications Return Information ORDERING/RETURN Information Ordering InformationPage Plasma Thawing Water Bath System Warranty