Genesis Advanced Technologies 2002, 2006 user manual Unit does not rock

Page 16

2)The ADD WATER light is on and there is sufficient water in the bath.

With the unit unplugged and raised out of the water, check that there is no obstruction around the float switch in the shroud unit. The float should move freely up and down the stem approximately half-inch.

3)The FAULT light is on.

Check the water temperature. If the water temperature is above 58 °C, the FAULT light will light and the unit will not be operable.

If this temperature is reached, it may indicate a unit malfunction.

Contact GenesisBPS, LLC for further assistance.

4)Unit does not heat. HEATING light is on but actual water temperature is not increasing.

Remove rear cover and check the heating element for continuity. Replace if necessary. This maintenance should only be performed by a trained technician or GenesisBPS, LLC. Contact it or your Biomedical Department.

5)Unit does not rock

Check that the ADD WATER or FAULT lights are not illuminated.

6)Bath moves to side when rocking or the rocking motion is not smooth.

Ensure that the bath is positioned on the white roller.

The unit must be operated on a smooth and level surface. Check the level of the work surface and make adjustments if necessary.

16

Image 16
Contents Genesis 2002TM Genesis 2006TM Genesis 2006 Shown GenesisBPS, LLC Contents Unpacking To Thaw Blood ProductsPage Page Page To Remove a Thawed BAG To Maintain the Device in the READY-TO- Thaw StatusPage Page Control Panel and Indicators Genesis 2002TM Control Panel Genesis 2006TM Control Panel Control Keys Indicators Unit does not rock Pump does not circulate the water Special Features Unique SafeguardsSpecifications ORDERING/RETURN Information Ordering Information Return InformationPage Warranty Plasma Thawing Water Bath System