1.UNPACKING
1.1The following is supplied with your GENESIS 2002/2006 unit.
Control Unit Water Bath
Thawing Bag Dispenser
Plasma Overwrap Thawing Bags (1000/Box) Plasma Bag Holder
LABath CLEANTM Reference Thermometer Thawing Bag Weights User's Manual
Optional Accessories :
Additional Bag Weights
The above items may be purchased from :
GenesisBPS, LLC
2.TO THAW BLOOD PRODUCTS
2.1 Place the unit on a flat, sturdy surface.
2.2Fill bath with water up to the slotted line on the front of the stainless steel shroud. This level will ensure proper thawing operation. No bags should be in the bath and the unit should be placed in the down position when filling the bath.
Check water level daily. If the water level is too low, the GENESIS 2002/2006 low water level sensing switch will make the unit inoperable (when the unit is plugged into the power supply). An alarm will sound and the ADD WATER
5