Genesis Advanced Technologies 2006, 2002 user manual Pump does not circulate the water

Page 17

7)Pump does not circulate the water.

Debris may be in the pump impeller housing. Raise the shroud unit and lock in the up position. Inspect pump for debris and remove with tweezers or similar instrument.

8)HEATING light is flashing.

This is normal operation. At approximately one degree below the set temperature, the microprocessor begins to reduce the amount of heat generated and the HEATING light begins to flash.

9)Temperature display at WATER TEMP deviates from thermometer reading.

Run the unit for more than 30 minutes.

Read thermometer temperature while the thermometer remains in water. Remember the temperature reading.

Press the POWER key to turn the unit power off.

Press the SET TEMP key and the POWER key at the same time. Power is turned on again but the WATER TEMP display shows an arbitrary number.

Each time the [5] key is pressed, the number displayed

in WATER TEMP window increases by 0.1. Each time the [0] key is pressed, it decreases by 0.1.

Press either the [5] key or the [0] key until the WATER TEMP shows same number as the thermometer reading. This adjustment should be done in one minute from when you have read the thermometer.

Press the POWER key to turn power off.

Press the POWER key to turn power on again and check calibration has been done correctly.

17

Image 17
Contents Genesis 2002TM Genesis 2006TM Genesis 2006 Shown GenesisBPS, LLC Contents To Thaw Blood Products UnpackingPage Page Page To Maintain the Device in the READY-TO- Thaw Status To Remove a Thawed BAGPage Page Control Panel and Indicators Genesis 2002TM Control Panel Genesis 2006TM Control Panel Control Keys Indicators Unit does not rock Pump does not circulate the water Unique Safeguards Special FeaturesSpecifications Return Information ORDERING/RETURN Information Ordering InformationPage Plasma Thawing Water Bath System Warranty