Genesis Advanced Technologies 2006, 2002 user manual Indicators

Page 15

INDICATORS

HEATING ....... This red indicator is lit when the heating

element is actually drawing current from the circuit and heating the water. When it reaches set temperature, the heating indicator light shuts off.

FAULT ............ If the water temperature exceeds 58 °C,

an alarm will sound and the red fault light will be illuminated. This may indicate a unit malfunction.

ADD WATER .... This red indicator will light if there is

insufficient water in the bath. The unit will not operate. Add water to correct this condition. This indicator will light also when the shroud unit is lifted in up position.

7.TROUBLESHOOTING GUIDE PROBLEM and ACTION

1)The indicator lights and displays do not come on when the POWER key is pressed.

Check the POWER SWITCH on the left-hand side.

Check the outlet for 110/120 VAC 60 Hz.

Check the fuse in the unit. Replace the fuse if necessary.

Ensure that the POWER key is pressed for at least one second.

Check that there is sufficient water in the bath and that

the ADD WATER light is off.

Check that the FAULT light is off.

15

Image 15
Contents Genesis 2002TM Genesis 2006TM Genesis 2006 Shown GenesisBPS, LLC Contents To Thaw Blood Products UnpackingPage Page Page To Maintain the Device in the READY-TO- Thaw Status To Remove a Thawed BAGPage Page Control Panel and Indicators Genesis 2002TM Control Panel Genesis 2006TM Control Panel Control Keys Indicators Unit does not rock Pump does not circulate the water Unique Safeguards Special FeaturesSpecifications Return Information ORDERING/RETURN Information Ordering InformationPage Plasma Thawing Water Bath System Warranty