Operating the lamp | Observe the safety notes on page 7. |
| Observe the notes with respect to the use of the trans- |
| former on page 38. |
Connect the cable between the lamp housing and the transformer (page 38).
Switch on the transformer and use the rotary knob to adjust the brightness (page 39).
Hold the lamp housing and release the clamping screw.
Align the light spot with the object.
Retighten the clamping screw.
Swing in the
Enlarge or reduce the size of the light
spot by turning the socket of the bulb.
Changing | Disconnect the lamp cable from the transformer. |
the bulb |
|
Never touch the bulb when it is hot.
You may sustain burns.
•You can separate the lamp housing and the bulb socket. Line up the arrow head on the mount with its shaft on the housing.
•Do not touch the 6 V/10 W halogen bulb because finger marks can cause it to crack when heated. Remove finger marks with a cloth dipped in alcohol.
Turn the socket to the separation position.
Pull the socket forcibly out of the housing.
Hold the new bulb in a cloth and push it into position.
Replace the socket in the housing. The socket clicks into position.
Leica M Stereomicroscopes – Illuminations | 41 |