Emerson Insertion Installations, Model 399VP Shown in LOW Flow Cell Assembly PN

Page 10

MODEL 399VP pH/ORP SENSOR

SECTION 2.0

 

INSTALLATION

The see-through flow cell is perfect for processes where flow regulation is desired. Tubing (1/4 inch) is needed for connection to the process or sample stream.

FIGURE 2-3. MODEL 399VP SHOWN IN LOW FLOW CELL ASSEMBLY (PN 23728-00)

2.4 INSERTION INSTALLATIONS.

The Model 399VP sensor has forward and rear facing 1 in. MNPT process connections for pipeline installations (see Figure 2-4 or 2-5) and flange connections. The Model 399VP is also suitable for side-of-tank installation. The Model 399VP must be mounted at least 10° above the horizontal (see Figure 2-4).

NOTE

Large pipe wrenches must not be used to tighten the sensor into a flange or other type of mounting.

FIGURE 2-4. MODEL 399VP SHOWN IN VARIOUS INSERTION INSTALLATIONS

Model 399VP can be mounted using front or back threads to install sensor with flange into process. Flange is supplied by others.

6

Image 10
Contents Combination pH/ORP Sensor Hazardous Area Installation Essential InstructionsTable of Contents Model 399VP pH/ORP SensorList of Figures List of TablesPhysical Specifications Section Description and SpecificationsFeatures and Applications Graph A. Pressure/Temperature Limitations For Model 399VPMounting Accessories optional Ordering InformationMounting Section InstallationUnpacking and Inspection Mounting GuidelinesSensor Dimensions FLOW-THROUGH Installations Model 399VP Shown in Various Flow Through InstallationsInsertion Installations Model 399VP Shown in LOW Flow Cell Assembly PNSubmersion Installations Adapter PNSensor Shown in Handrail Mounting Assembly PN JET Spray Cleaner PN Used with Model 399VP Section Wiring SCLWire and Connector Pin Functions Model 54 Wiring Model 81 Wiring through Remote Junction Box 13. Model 1054 Wiring 17. Model SCL-P/Q Wiring 41055107 41055108 23. Preparation of Extension Cable PN Sensor Preparation Recommended two-point buffer calibration procedureSection START-UP and Calibration Recommended pH Sensor Standardization268 264 260 Automatic Temperature Compensator Section MaintenanceGeneral PH Electrode Cleaning ORP Platinum Electrode CheckSection Troubleshooting Problem Probable Cause RemedySection Return of Material GeneralReturn of Materials Request Sensor or Circuit Board onlyEUROPE, Middle EAST, and Africa Americas HeadquartersASIA-PACIFIC Additional Sales OfficesWarranty Specifications subject to change without notice