Emerson Combination pH/ORP Sensor, 399VP instruction manual Submersion Installations, Adapter PN

Page 11

MODEL 399VP pH/ORP SENSOR

SECTION 2.0

 

INSTALLATION

FIGURE 2-5. MODEL 399VP WITH INSERTION

ADAPTER (PN 23242-02)

The Insertion Adapter mounts Model 399VP into a 1-1/2 inch NPT process pipe. Unscrewing the hex nut allows easy sensor removal without twisting the sensor cable.

2.5 SUBMERSION INSTALLATIONS.

The Model 399VP Sensor has a 1 in. MNPT process connection at the back of the sensor (see Figure 2-1). Utilizing a standard 1 in. union (such as PN 9330022), the sensor may be mounted to a 1 in. SCH 80 CPVC or PVDF stand- pipe. The sensor can also be mounted in the Handrail Mounting Assembly (PN 11275-01; see Figure 2-6). Tapered pipe threads in plastic tend to loosen after installation. It is therefore recommended that the tightness of the con- nection be checked frequently to ensure that no loosening has occurred. To prevent rain water or condensation from running into the sensor, a weatherproof junction box is recommended (Rosemount Analytical PN 22719-02; see Figure 2-8). The sensor cable must be run through a protective conduit for isolation from electrical interference or physical abuse from the process. The sensor should be installed within 80° of vertical, with the electrode facing down. The sensor's cable should not be run with power or control wiring.

NOTE

LARGE PIPE WRENCHES MUST NOT BE USED TO TIGHTEN THE SENSOR INTO MOUNTING ACCESSORIES.

When the Model 399VP is installed using a 1 in. standpipe, the sensor can be cleaned while still submersed in the process by using a jet spray cleaner. The jet spray cleaner (PN 12707-00; see Figure 2-7) can spray water or air directly onto the sensor.

7

Image 11
Contents Combination pH/ORP Sensor Essential Instructions Hazardous Area InstallationModel 399VP pH/ORP Sensor Table of ContentsList of Tables List of FiguresGraph A. Pressure/Temperature Limitations For Model 399VP Section Description and SpecificationsFeatures and Applications Physical SpecificationsOrdering Information Mounting Accessories optionalMounting Guidelines Section InstallationUnpacking and Inspection MountingSensor Dimensions Model 399VP Shown in Various Flow Through Installations FLOW-THROUGH InstallationsModel 399VP Shown in LOW Flow Cell Assembly PN Insertion InstallationsAdapter PN Submersion InstallationsSensor Shown in Handrail Mounting Assembly PN JET Spray Cleaner PN Used with Model 399VP SCL Section WiringWire and Connector Pin Functions Model 54 Wiring Model 81 Wiring through Remote Junction Box 13. Model 1054 Wiring 17. Model SCL-P/Q Wiring 41055107 41055108 23. Preparation of Extension Cable PN Recommended pH Sensor Standardization Recommended two-point buffer calibration procedureSection START-UP and Calibration Sensor Preparation268 264 260 General Section MaintenanceAutomatic Temperature Compensator ORP Platinum Electrode Check PH Electrode CleaningProblem Probable Cause Remedy Section TroubleshootingGeneral Section Return of MaterialSensor or Circuit Board only Return of Materials RequestAdditional Sales Offices Americas HeadquartersASIA-PACIFIC EUROPE, Middle EAST, and AfricaWarranty Specifications subject to change without notice