Emerson Combination pH/ORP Sensor Section Description and Specifications, Physical Specifications

Page 5

MODEL 399VP pH/ORP SENSOR

SECTION 1.0

 

DESCRIPTION AND SPECIFICATIONS

SECTION 1.0.

DESCRIPTION AND SPECIFICATIONS

1.1FEATURES AND APPLICATIONS

1.2PHYSICAL SPECIFICATIONS

1.3ORDERING INFORMATION

1.1 FEATURES AND APPLICATIONS

The Rosemount Analytical Model 399VP Sensor meas- ures the pH or the Oxidation Reduction Potential (ORP) of aqueous solutions in pipelines, open tanks, or ponds. It is suitable for virtually all applications where a low cost sensor is required. The combination electrode features a ceramic junction constructed in an annular design around the pH/ORP sensitive membrane. The double or triple junction reference cell configuration is resistant to process solutions containing ammonia, chlorine, cyanides, sulfides, or other poisoning ions.

The glass electrode is housed in a molded Tefzel body and sealed with Viton1 O-rings to guard against process leakage. The cable end of the sensor is also sealed, eliminating cable shorts caused by exposure to moisture. This construction makes for a highly chemical resistant disposable sensor and does not require electrolyte replenishment.

The custom glass electrode is available with either a standard hemi bulb or an optional flat bulb to best meet your application needs. Flat glass is advanta- geous in abrasive or coating applications that etch or build up on the glass respectively. In coating applica- tions, such as slurries, the flat surface allows the process flow to act as a scrubbing agent to reduce coating and maintenance whereas in abrasive appli- cations pitting from silicates and other similar materi- als is minimized by the flat surface to provide longer life. The hemi bulb is ideal for general purpose use and for those processes requiring greater accuracy over the entire pH range

Installation is easily achieved through the wide variety of mounting configurations. The Model 399VP fea- tures 1 in. (MNPT) front and rear facing connections for insertion, submersion or flow through pH and ORP applications.

1.2 PHYSICAL SPECIFICATIONS

Materials of Construction: Tefzel, glass, ceramic and Viton (also, Platinum for ORP sensor)

Process Connections: 1 in. MNPT

Interconnecting Cable: None - must use mating VP cable

Measured Range: pH: 0-14*

ORP: –1500 -+1500mV *Percent Linearity over pH range:

 

Hemi Bulb

Flat Bulb

1-2 pH

94%

93%

2-12 pH

99%

98%

12-13 pH

97%

95%

 

 

 

Temperature Compensation: Automatic 0 to 85°C (32° to 185°F) (Temperature compensation is not required for 399 ORP when used with Models 1060, 1023 or 1181 ORP)

Maximum Pressure: 790 kPa abs (100 psig) at 65°C (Refer to Graph A below)

Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb)

1Registered trademarks of E. I. du Pont de Nemours and Company.

GRAPH A. Pressure/Temperature Limitations

for Model 399VP

1

Image 5
Contents Combination pH/ORP Sensor Essential Instructions Hazardous Area InstallationModel 399VP pH/ORP Sensor Table of ContentsList of Tables List of FiguresFeatures and Applications Section Description and SpecificationsPhysical Specifications Graph A. Pressure/Temperature Limitations For Model 399VPOrdering Information Mounting Accessories optionalUnpacking and Inspection Section InstallationMounting Mounting GuidelinesSensor Dimensions Model 399VP Shown in Various Flow Through Installations FLOW-THROUGH InstallationsModel 399VP Shown in LOW Flow Cell Assembly PN Insertion InstallationsAdapter PN Submersion InstallationsSensor Shown in Handrail Mounting Assembly PN JET Spray Cleaner PN Used with Model 399VP SCL Section WiringWire and Connector Pin Functions Model 54 Wiring Model 81 Wiring through Remote Junction Box 13. Model 1054 Wiring 17. Model SCL-P/Q Wiring 41055107 41055108 23. Preparation of Extension Cable PN Section START-UP and Calibration Recommended two-point buffer calibration procedureSensor Preparation Recommended pH Sensor Standardization268 264 260 General Section MaintenanceAutomatic Temperature Compensator ORP Platinum Electrode Check PH Electrode CleaningProblem Probable Cause Remedy Section TroubleshootingGeneral Section Return of MaterialSensor or Circuit Board only Return of Materials RequestASIA-PACIFIC Americas HeadquartersEUROPE, Middle EAST, and Africa Additional Sales OfficesWarranty Specifications subject to change without notice