Emerson Combination pH/ORP Sensor, 399VP Section Troubleshooting, Problem Probable Cause Remedy

Page 27

MODEL 399VP pH/ORP SENSOR

SECTION 6.0

 

TROUBLESHOOTING

SECTION 6.0.

TROUBLESHOOTING

TROUBLESHOOTING. The table below lists common problems, causes and remedies typically encountered in process measurement.

Problem

Probable Cause

 

Remedy

 

 

 

 

 

Meter reads off scale (display

T.C. element shorted.

 

Check T.C. element as instructed

reads overrange).

 

 

in Section 5.2 and replace

 

 

 

sensor if defective.

 

 

 

 

 

Sensor not in process. Sample

 

Make sure sensor is in process with

 

stream is low or air bubbles are

 

sufficient sample stream (refer to

 

present.

 

Section 2.0 for installation details).

 

 

 

 

 

Open glass electrode.

 

Replace sensor.

 

 

 

 

 

Reference element open - no contact.

 

Replace sensor.

 

 

 

 

Display reads between 3 and 6 pH,

Electrode cracked.

 

Replace sensor.

regardless of actual pH of solution

 

 

 

or sample.

 

 

 

 

 

 

 

Meter or display indication swings

T.C. element shorted.

 

Check T.C. element as instructed

or jumps widely in AUTO T.C. Mode.

 

 

in Section 5.2 and replace

 

 

 

sensor if defective.

 

 

 

 

Span between buffers extremely

T.C. element open.

 

Check T.C. element as instructed

short in AUTO T.C. Mode.

 

 

in Section 5.2 and replace sensor

 

 

 

if defective.

 

 

 

 

Sluggish or slow meter indication

Electrode coated.

 

Clean sensor as instructed in

for real changes in pH level.

 

 

Section 5.3. Replace sensor if

 

 

 

cracked.

 

 

 

 

 

Electrode defective.

 

Replace sensor.

 

 

 

 

Transmitter cannot be standardized.

Electrode coated or cracked.

 

Clean sensor as instructed in

 

 

 

Section 5.3. Replace sensor if

 

 

 

cracked.

 

 

 

 

Transmitter short spans between

Aged glass electrode or high

 

Replace sensor.

two different buffer values.

temperature exposure.

 

 

 

 

 

 

 

Electrode coated .

 

Clean sensor as instructed in

 

 

 

Section 5.3. Replace sensor if

 

 

 

cracked.

 

 

 

 

 

 

 

 

23

Image 27
Contents Combination pH/ORP Sensor Essential Instructions Hazardous Area InstallationModel 399VP pH/ORP Sensor Table of ContentsList of Tables List of FiguresGraph A. Pressure/Temperature Limitations For Model 399VP Section Description and SpecificationsFeatures and Applications Physical SpecificationsOrdering Information Mounting Accessories optionalMounting Guidelines Section InstallationUnpacking and Inspection MountingSensor Dimensions Model 399VP Shown in Various Flow Through Installations FLOW-THROUGH InstallationsModel 399VP Shown in LOW Flow Cell Assembly PN Insertion InstallationsAdapter PN Submersion InstallationsSensor Shown in Handrail Mounting Assembly PN JET Spray Cleaner PN Used with Model 399VP SCL Section WiringWire and Connector Pin Functions Model 54 Wiring Model 81 Wiring through Remote Junction Box 13. Model 1054 Wiring 17. Model SCL-P/Q Wiring 41055107 41055108 23. Preparation of Extension Cable PN Recommended pH Sensor Standardization Recommended two-point buffer calibration procedureSection START-UP and Calibration Sensor Preparation268 264 260 Section Maintenance Automatic Temperature CompensatorGeneral ORP Platinum Electrode Check PH Electrode CleaningProblem Probable Cause Remedy Section TroubleshootingGeneral Section Return of MaterialSensor or Circuit Board only Return of Materials RequestAdditional Sales Offices Americas HeadquartersASIA-PACIFIC EUROPE, Middle EAST, and AfricaWarranty Specifications subject to change without notice