Output On/Off Synchronization Appendix D

Synchronizing Multiple Mainframes

The output turn-on synchronization function can be used across multiple mainframes that have coupled output channels. Each mainframe that will be synchronized must have at least one coupled channel. Note that cross-frame synchronization must be enabled for any mainframe that contains synchronized output channels.

Procedure

1.Configure the output channels on each mainframe as described in steps 1 through 3 of the previous procedure.

2.This step is required if you have power modules with different minimum delay offsets. Specify a common delay offset for all of the synchronized output channels. This value must be the largest delay offset of all synchronized output channels regardless of the mainframe in which they are installed. This same value must be specified as the common delay offset for each mainframe.

Front Panel:

SCPI Command:

In the front panel menu of each

To specify the common delay offset

mainframe, select Output\Couple.

for each mainframe in seconds:

In the Delay offset field, enter the

OUTP:COUP:DOFF .051

 

delay offset value of the slowest

To return the delay offset of the

power module of all the mainframes

slowest power module in each

in milliseconds; then press Select.

mainframe (the maximum delay

The Max delay offset for this frame

offset) in seconds:

field displays the delay offset of the

OUTP:COUP:MAX:DOFF?

slowest power module in this frame.

 

3.Connect and configure the digital connector pins of the synchronized mainframes as described in the following section.

Digital Connections and Configuration

 

Only pins 4 through 7 can be configured as synchronization pins. You cannot

NOTE

 

configure more than one ON couple and one OFF couple pin per mainframe.

 

The polarity of the pins is not programmable.

 

 

 

The digital connector pins of the synchronized mainframes that

 

contain coupled channels must be connected together as shown in

 

the following figure. In this example, pin 6 will be configured as the

 

output On control. Pin 7 will be configured as the output Off control.

 

The ground or Common pins also need to be connected together.

Series N6700 User’s Guide

103

Image 103
Contents Agilent Technologies Low-Profile Modular Power System Agilent Technologies, Inc Legal NoticesGeneral Safety NoticesThis Book Contents 104 Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A100 103Essential operating features of the power system Modular Power System MPSReferred to as MPS and power system throughout this manual Output Features Agilent N6700 Modular Power System At a GlanceModel Differences System FeaturesRear Panel At a Glance Front Panel At a GlanceInterface status IndicatorsFront Panel Display At a Glance Single-channel viewOperating status Front Panel Keys At a GlanceNumber keys OCP Front Panel Menu ReferenceUSB Subsystem Commands Scpi Command SummaryScpi Command Description FUNCtion VOLTage CURRent, @chanlist SWEep Scpi CommandALL These settings are set by the *RST Reset command Common CommandsRST Settings Command DescriptionInstallation Options General InformationModels Items SuppliedInspecting the Unit Installing the UnitSafety Considerations Rack InstallationInstallation Hz Operation Bench InstallationCleaning Channel NumberSnap-On Ferrite Core Connecting the Line CordInstallation Power cord, resulting in fireTerminals in any other way Connecting the OutputsScrews securely tightened Terminals. Sense jumpers are provided for local sensingAWG Wire SizeResponse Time with an External Capacitor Multiple LoadsPositive and Negative Voltages N6700 Modular Power System Scope Drop in the sense leads can degrade the voltage regulation Remote Sense ConnectionsOutput Noise Considerations Parallel ConnectionsOpen Sense Leads Over-voltage Protection ConsiderationsDiscussed in under System-Related Operations When programming via the front panel or using Scpi commandsGrouping the Outputs Firmware revision B.00.00 and upTerminal may be more than 240 VDC from chassis ground Effect on SpecificationsRecovery Time Series ConnectionsDesired operating voltage Setting the OutputsEffect on Specifications Menu structure is found in Chapter Getting StartedMethod 1 Use the Navigation and Arrow Keys Entering an Output Voltage SettingTurning the Unit On Selecting an Output ChannelMethod 2 Use the Current key to enter a value Entering a Current Limit SettingUse the On/Off key to enable the output Enabling the OutputMenu Using the Front Panel MenuPress the Menu key to access the front panel command menu Set the Over-Voltage ProtectionEnter Channel Exiting the Command MenuMeter Back There are two ways to exit the command menu Case of TroubleAutomation-Ready CD that is shipped with your product Connecting to the InterfacesCause the unit to reset and require operator intervention Gpib InterfaceFollowing figure illustrates a typical USB interface system USB-enabled instrument to the Universal Serial Bus USB.USB Interface LAN InterfaceConnecting to a Private LAN Connecting to a Site LANWill come on when the LAN port has been configured Release the previous settingsFactory-shipped instrument LAN settings are configured to Dhcp server. You can leave these settings as they are. MostTake effect Viewing the Currently Active LAN SettingsConfiguring the LAN Parameters LAN ParametersDomain NameLAN Communication ResetReduced Using the Web ServerUsing Sockets Using TelnetTelnet connections to be made Restoring the Non-volatile Factory Settings Enable/Disable the USB, LAN, and Web ServerSecuring the Interfaces Other factory-shipped non-volatile settings Factory-shipped non-volatile LAN settingsOperating the Power System Set the Voltage Slew Rate Programming the OutputSelect an Output Channel Set the Output VoltageOutpdelrise .1,@2 Enable the OutputSet the Output Current Sequence Multiple OutputsOutprelpol REV,@1 Front Panel Scpi Command Select Output\PolarityProgram the Output Relays OutpSynchronizing Output Steps Enable the Output to Respond to Trigger CommandsFront Panel Scpi Command Select Transient\Mode Front Panel Scpi Command Select Transient\StepInitiate the Transient Trigger System Select the Transient Trigger SourceScroll to and select Abort Generate Trigger Out SignalsFront Panel Scpi Command Select Transient\Control Trigger the OutputSimultaneous Voltage and Current Measurements Making MeasurementsMeasurement Ranges Enter a value in the OVP level box To 10 Press Select Using the Protection FunctionsFront Panel Scpi Command Select Protect\OVP Set the Over-Current ProtectionClear Output Protection Functions Couple Output ProtectionInstrument State Storage System-Related OperationsSelf-Test Instrument IdentificationSystreb Front Panel Scpi Command Select System\GroupsOutput Groups SystgrodelallLockout Front Panel Scpi Command Select System\Preferences\LockFront Panel Scpi Command Select System\Preferences\Keys Front Panel KeysView Front Panel DisplayScreen Saver ContrastAll models Refer to , Model Differences When the power system is turned offProgramming High-Speed Test Extensions List FunctionVoltmode LIST, @1 Front Panel Scpi Command Select Transient\List\ConfigFront Panel Scpi Command Select Transient\List\Pace Program an Output Pulse or Pulse TrainInittran @1 Front Panel Scpi Command Select Transient\List\RepeatFront Panel Scpi Command Select Transient\TrigSource Trigtransour BUS, @1Select the List Step number Program an Arbitrary ListListstep AUTO, @1 Front Panel Scpi Command Select the Transient\List\RepeatProgramming the Digitizer Adjust the Measurement Sample RateFront Panel Scpi Command Select Measure\Sweep Digitizer FunctionAcquisition trigger, use Specify a Window FunctionTo offset the measurement on Retrieve Measurement Array Data Front Panel Scpi Command Select Measure\WindowFront Panel Scpi Command Not Available Synchronizing Digitizer MeasurementsTransientn Select the Measurement Trigger SourceSensfunc VOLT,@14 Sensfunc CURR,@14Trigacq @1 Front Panel Scpi Command Select Measure\ControlInitiate the Measurement Trigger System Trigger the MeasurementAppendix a Specifications Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A Performance SpecificationsSupplemental Characteristics Down-programming Time with Capacitive load Up-programming Time with full resistive loadUp-programming Settling Time with full resistive load Down-programming Settling Time with no loadOutput Impedance Graphs Autoranging CharacteristicModel N6753A, CV Mode, @20V, 15A 0125 Magnitude Agilent Models N6731B N6736B and N6741B N6746B Load Transient Recovery TimeCommon Mode Noise Temperature Coefficient per CFrom 20 Hz 20 MHz from either output to chassis Agilent Models N6773A N6776A 20 mA Agilent N6700B, N6701A, N6702A MPS Mainframes AC Input DimensionsNet Weight Environmental Conditions432.5 mm 17.03 482.6 mm 19.00 425.45 mm Outline DiagramAppendix B Using the Digital Port Digital Control Port Select System\IO\DigPort\Pinn Configuring the Digital Control PortBi-directional Digital I/O Pin BitDigital Input Front Panel Scpi Command Select System\IO\DigPort\PinnExternal Trigger Inhibit Input Fault OutputFront Panel Scpi Command Select System\IO\DigPort\Pin1 Front Panel Scpi Command Select System\IO\DigPort\Pin3Clearing a System Protection Fault Fault/Inhibit System ProtectionMainframes Rating of the mainframe This chapter discusses the power allocation functionFor the majority of Agilent N6700 Modular Power System Power rating of Agilent N6700 mainframes is as followsMust reset the power limit to its maximum rated value Power Limit OperationMainframe Power Limit Module Power LimitTo set a power limit on output Module Power AllocationPower limit function will not activate Page Appendix D Output On/Off Synchronization Power Modules Options and Mode Minimum Synchronizing Output Turn-on DelaysTutorial Outpcoup OFF ProcedureEnable the Synchronization Function Outpcoup onSpecify the Common Delay Offset Specify the Turn-On Delays for each Output ChannelSpecify which Output Channels will be Synchronized Synchronizing Multiple Mainframes Digital Connections and ConfigurationDIGPIN7FUNC Offc OperationSystem\IO\DigPort\Pin6 DIGPIN6FUNC ONCIndex List Inhibit inputInstrument 11, 42Remote interface SpecificationsSystem protection Declaration of Conformity Manual Updates