Agilent Technologies N6700 manual Bench Installation, Channel Number, Hz Operation, Cleaning

Page 23

Installation 2

Bench Installation

Do not block the air intake and exhaust at the sides, or the exhaust at the rear of the unit. Refer to the outline diagram in Appendix A. Minimum clearances for bench operation are 2 inches (51 mm) along the sides and back.

Channel Number

 

The channel number of a power module is determined by the location

 

of that module in the mainframe. When viewed from the rear, the

 

module next to the GPIB connector is always output channel one.

 

Numbering continues sequentially to the left, from one to four.

 

If there are less than four modules, channel numbering corresponds

 

to the actual number of installed power modules. Unused channel

 

slots must contain filler modules to ensure proper airflow for cooling.

 

Double-wide power modules are assigned the channel number of the

 

lowest numbered slot in which is installed. For example, if a double-

 

wide module is installed in slots 3 and 4, it is assigned channel

 

number 3.

 

Power modules that are connected in parallel and have been configured or

NOTE

grouped to act as a single, higher-power channel are addressed using the

 

 

channel number of the lowest channel in the group.

 

 

400 Hz Operation

Redundant Ground Requirement

At 400 Hz AC input operation, the leakage current of the unit exceeds

3.5mA. This requires the installation of a permanent, redundant ground from the instrument chassis to earth ground. This ensures that ground will always be connected and that any leakage current will be diverted to ground. Refer to the Service Guide for installation instructions.

Power Factor

At 400 Hz AC input operation, the unit’s power factor is affected as follows:

Under full load at 400 Hz, power factor drops from 0.99 (@120

VAC) to as low as 0.76 (@ 265 VAC).

Power factor degrades further under no load conditions.

Cleaning

WARNING SHOCK HAZARD To prevent electric shock, unplug the unit before cleaning.

Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.

Series N6700 User’s Guide

23

Image 23
Contents Agilent Technologies Low-Profile Modular Power System Agilent Technologies, Inc Legal NoticesGeneral Safety NoticesThis Book Contents 104 Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A100 103Referred to as MPS and power system throughout this manual Modular Power System MPSEssential operating features of the power system Output Features Agilent N6700 Modular Power System At a GlanceModel Differences System FeaturesRear Panel At a Glance Front Panel At a GlanceInterface status IndicatorsFront Panel Display At a Glance Single-channel viewNumber keys Front Panel Keys At a GlanceOperating status OCP Front Panel Menu ReferenceUSB Scpi Command Description Scpi Command SummarySubsystem Commands FUNCtion VOLTage CURRent, @chanlist SWEep Scpi CommandALL These settings are set by the *RST Reset command Common CommandsRST Settings Command DescriptionInstallation Options General Information Models Items SuppliedInspecting the Unit Installing the UnitSafety Considerations Rack InstallationInstallation Hz Operation Bench InstallationCleaning Channel NumberSnap-On Ferrite Core Connecting the Line CordInstallation Power cord, resulting in fireTerminals in any other way Connecting the OutputsScrews securely tightened Terminals. Sense jumpers are provided for local sensingAWG Wire SizePositive and Negative Voltages Multiple LoadsResponse Time with an External Capacitor N6700 Modular Power System Scope Drop in the sense leads can degrade the voltage regulation Remote Sense ConnectionsOutput Noise Considerations Parallel ConnectionsOpen Sense Leads Over-voltage Protection ConsiderationsDiscussed in under System-Related Operations When programming via the front panel or using Scpi commandsGrouping the Outputs Firmware revision B.00.00 and upTerminal may be more than 240 VDC from chassis ground Effect on SpecificationsRecovery Time Series ConnectionsDesired operating voltage Setting the OutputsEffect on Specifications Menu structure is found in Chapter Getting StartedMethod 1 Use the Navigation and Arrow Keys Entering an Output Voltage SettingTurning the Unit On Selecting an Output ChannelMethod 2 Use the Current key to enter a value Entering a Current Limit SettingUse the On/Off key to enable the output Enabling the OutputMenu Using the Front Panel MenuPress the Menu key to access the front panel command menu Set the Over-Voltage ProtectionEnter Channel Exiting the Command MenuMeter Back There are two ways to exit the command menu Case of TroubleAutomation-Ready CD that is shipped with your product Connecting to the InterfacesCause the unit to reset and require operator intervention Gpib InterfaceFollowing figure illustrates a typical USB interface system USB-enabled instrument to the Universal Serial Bus USB.USB Interface LAN InterfaceConnecting to a Private LAN Connecting to a Site LANWill come on when the LAN port has been configured Release the previous settingsFactory-shipped instrument LAN settings are configured to Dhcp server. You can leave these settings as they are. MostTake effect Viewing the Currently Active LAN SettingsConfiguring the LAN Parameters LAN ParametersDomain NameLAN Communication ResetReduced Using the Web ServerTelnet connections to be made Using TelnetUsing Sockets Securing the Interfaces Enable/Disable the USB, LAN, and Web ServerRestoring the Non-volatile Factory Settings Other factory-shipped non-volatile settings Factory-shipped non-volatile LAN settingsOperating the Power System Set the Voltage Slew Rate Programming the OutputSelect an Output Channel Set the Output VoltageOutpdelrise .1,@2 Enable the OutputSet the Output Current Sequence Multiple OutputsOutprelpol REV,@1 Front Panel Scpi Command Select Output\PolarityProgram the Output Relays OutpSynchronizing Output Steps Enable the Output to Respond to Trigger CommandsFront Panel Scpi Command Select Transient\Mode Front Panel Scpi Command Select Transient\StepInitiate the Transient Trigger System Select the Transient Trigger SourceScroll to and select Abort Generate Trigger Out SignalsFront Panel Scpi Command Select Transient\Control Trigger the OutputMeasurement Ranges Making MeasurementsSimultaneous Voltage and Current Measurements Enter a value in the OVP level box To 10 Press Select Using the Protection FunctionsFront Panel Scpi Command Select Protect\OVP Set the Over-Current ProtectionClear Output Protection Functions Couple Output ProtectionInstrument State Storage System-Related OperationsSelf-Test Instrument IdentificationSystreb Front Panel Scpi Command Select System\GroupsOutput Groups SystgrodelallLockout Front Panel Scpi Command Select System\Preferences\LockFront Panel Scpi Command Select System\Preferences\Keys Front Panel KeysView Front Panel DisplayScreen Saver ContrastAll models Refer to , Model Differences When the power system is turned offProgramming High-Speed Test Extensions List FunctionVoltmode LIST, @1 Front Panel Scpi Command Select Transient\List\ConfigFront Panel Scpi Command Select Transient\List\Pace Program an Output Pulse or Pulse TrainInittran @1 Front Panel Scpi Command Select Transient\List\RepeatFront Panel Scpi Command Select Transient\TrigSource Trigtransour BUS, @1Select the List Step number Program an Arbitrary ListListstep AUTO, @1 Front Panel Scpi Command Select the Transient\List\RepeatProgramming the Digitizer Adjust the Measurement Sample RateFront Panel Scpi Command Select Measure\Sweep Digitizer FunctionTo offset the measurement on Specify a Window FunctionAcquisition trigger, use Retrieve Measurement Array Data Front Panel Scpi Command Select Measure\WindowFront Panel Scpi Command Not Available Synchronizing Digitizer MeasurementsTransientn Select the Measurement Trigger SourceSensfunc VOLT,@14 Sensfunc CURR,@14Trigacq @1 Front Panel Scpi Command Select Measure\ControlInitiate the Measurement Trigger System Trigger the MeasurementAppendix a Specifications Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A Performance SpecificationsSupplemental Characteristics Down-programming Time with Capacitive load Up-programming Time with full resistive loadUp-programming Settling Time with full resistive load Down-programming Settling Time with no loadOutput Impedance Graphs Autoranging CharacteristicModel N6753A, CV Mode, @20V, 15A 0125 Magnitude Agilent Models N6731B N6736B and N6741B N6746B Load Transient Recovery TimeFrom 20 Hz 20 MHz from either output to chassis Temperature Coefficient per CCommon Mode Noise Agilent Models N6773A N6776A 20 mA Agilent N6700B, N6701A, N6702A MPS Mainframes AC Input DimensionsNet Weight Environmental Conditions432.5 mm 17.03 482.6 mm 19.00 425.45 mm Outline DiagramAppendix B Using the Digital Port Digital Control Port Select System\IO\DigPort\Pinn Configuring the Digital Control PortBi-directional Digital I/O Pin BitExternal Trigger Front Panel Scpi Command Select System\IO\DigPort\PinnDigital Input Inhibit Input Fault OutputFront Panel Scpi Command Select System\IO\DigPort\Pin1 Front Panel Scpi Command Select System\IO\DigPort\Pin3Mainframes Fault/Inhibit System ProtectionClearing a System Protection Fault Rating of the mainframe This chapter discusses the power allocation functionFor the majority of Agilent N6700 Modular Power System Power rating of Agilent N6700 mainframes is as followsMust reset the power limit to its maximum rated value Power Limit OperationMainframe Power Limit Module Power LimitPower limit function will not activate Module Power AllocationTo set a power limit on output Page Appendix D Output On/Off Synchronization Tutorial Synchronizing Output Turn-on DelaysPower Modules Options and Mode Minimum Outpcoup OFF ProcedureEnable the Synchronization Function Outpcoup onSpecify which Output Channels will be Synchronized Specify the Turn-On Delays for each Output ChannelSpecify the Common Delay Offset Synchronizing Multiple Mainframes Digital Connections and ConfigurationDIGPIN7FUNC Offc OperationSystem\IO\DigPort\Pin6 DIGPIN6FUNC ONCIndex List Inhibit inputInstrument 11, 42System protection SpecificationsRemote interface Declaration of Conformity Manual Updates