Agilent Technologies N6700 Set the Output Current, Enable the Output, Sequence Multiple Outputs

Page 53

 

 

Operating the Power System 4

Set the Output Current

 

 

 

 

 

Front Panel:

SCPI Command:

 

Press the Current key.

To set output 1 to 1 A:

 

Enter a value and press Select.

CURR 1,(@1)

 

 

 

 

To set all outputs to 2 A:

 

 

CURR 2,(@1:4)

 

 

 

For models with multiple ranges, you can select a lower range if you need better output resolution.

Front Panel:

SCPI Command:

Press the Current key.

Select a lower range and press Select.

To select the lower range, program a value that falls within the range:

CURR:RANG 1,(@1)

Enable the Output

Front Panel:

SCPI Command:

Press the On/Off key.

To enable/disable ALL outputs using the On/Off key, select System\Preferences\Keys. Check On/Off affects all channels.

The All indicator will be on.

To enable only output 1: OUTP ON,(@1)

To enable outputs 1-4:

OUTP ON,(@1:4)

Because of internal circuit start-up procedures and any installed relay options, output on may take between 35 and 50 milliseconds to complete its function. Conversely, output off may take between 20 and 25 milliseconds to complete its function.

To mitigate these built-in delays, you can program the output to zero volts rather than using the output on/off function.

Sequence Multiple Outputs

Turn-on and turn-off delays control the power-up and power-down sequencing of the output channels in relation to each other.

Front Panel:

SCPI Command:

Press the Channel key to select an

To program a 50 millisecond turn-on

output. Then select Output\Delay.

delay for output 1 and a 100

Select either Turn-on or Turn-off.

millisecond turn-on delay for output 2:

Enter a delay in seconds, then

OUTP:DEL:RISE .05,(@1)

press Select.

OUTP:DEL:RISE .1,(@2)

Select System\Preferences\Keys.

To program a 200 millisecond turn-off

Check On/Off affects all channels.

delay for outputs 3 and 4:

 

OUTP:DEL:FALL .2,(@3,4)

 

 

Series N6700 User’s Guide

53

Image 53
Contents Agilent Technologies Low-Profile Modular Power System Agilent Technologies, Inc Legal NoticesGeneral Safety NoticesThis Book Contents 100 Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A103 104Referred to as MPS and power system throughout this manual Modular Power System MPSEssential operating features of the power system Output Features Agilent N6700 Modular Power System At a GlanceModel Differences System FeaturesRear Panel At a Glance Front Panel At a GlanceFront Panel Display At a Glance IndicatorsSingle-channel view Interface statusNumber keys Front Panel Keys At a GlanceOperating status OCP Front Panel Menu ReferenceUSB Scpi Command Description Scpi Command SummarySubsystem Commands FUNCtion VOLTage CURRent, @chanlist SWEep Scpi CommandALL RST Settings Common CommandsCommand Description These settings are set by the *RST Reset commandInstallation Models General InformationItems Supplied OptionsSafety Considerations Installing the UnitRack Installation Inspecting the UnitInstallation Cleaning Bench InstallationChannel Number Hz OperationInstallation Connecting the Line CordPower cord, resulting in fire Snap-On Ferrite CoreScrews securely tightened Connecting the OutputsTerminals. Sense jumpers are provided for local sensing Terminals in any other wayAWG Wire SizePositive and Negative Voltages Multiple LoadsResponse Time with an External Capacitor N6700 Modular Power System Scope Drop in the sense leads can degrade the voltage regulation Remote Sense ConnectionsOpen Sense Leads Parallel ConnectionsOver-voltage Protection Considerations Output Noise ConsiderationsGrouping the Outputs When programming via the front panel or using Scpi commandsFirmware revision B.00.00 and up Discussed in under System-Related OperationsRecovery Time Effect on SpecificationsSeries Connections Terminal may be more than 240 VDC from chassis groundDesired operating voltage Setting the OutputsEffect on Specifications Menu structure is found in Chapter Getting StartedTurning the Unit On Entering an Output Voltage SettingSelecting an Output Channel Method 1 Use the Navigation and Arrow KeysUse the On/Off key to enable the output Entering a Current Limit SettingEnabling the Output Method 2 Use the Current key to enter a valuePress the Menu key to access the front panel command menu Using the Front Panel MenuSet the Over-Voltage Protection MenuMeter Back There are two ways to exit the command menu Exiting the Command MenuCase of Trouble Enter ChannelCause the unit to reset and require operator intervention Connecting to the InterfacesGpib Interface Automation-Ready CD that is shipped with your productUSB Interface USB-enabled instrument to the Universal Serial Bus USB.LAN Interface Following figure illustrates a typical USB interface systemConnecting to a Private LAN Connecting to a Site LANFactory-shipped instrument LAN settings are configured to Release the previous settingsDhcp server. You can leave these settings as they are. Most Will come on when the LAN port has been configuredConfiguring the LAN Parameters Viewing the Currently Active LAN SettingsLAN Parameters Take effectDomain NameLAN Communication ResetReduced Using the Web ServerTelnet connections to be made Using TelnetUsing Sockets Securing the Interfaces Enable/Disable the USB, LAN, and Web ServerRestoring the Non-volatile Factory Settings Other factory-shipped non-volatile settings Factory-shipped non-volatile LAN settingsOperating the Power System Select an Output Channel Programming the OutputSet the Output Voltage Set the Voltage Slew RateSet the Output Current Enable the OutputSequence Multiple Outputs Outpdelrise .1,@2Program the Output Relays Front Panel Scpi Command Select Output\PolarityOutp Outprelpol REV,@1Front Panel Scpi Command Select Transient\Mode Enable the Output to Respond to Trigger CommandsFront Panel Scpi Command Select Transient\Step Synchronizing Output StepsInitiate the Transient Trigger System Select the Transient Trigger SourceFront Panel Scpi Command Select Transient\Control Generate Trigger Out SignalsTrigger the Output Scroll to and select AbortMeasurement Ranges Making MeasurementsSimultaneous Voltage and Current Measurements Front Panel Scpi Command Select Protect\OVP Using the Protection FunctionsSet the Over-Current Protection Enter a value in the OVP level box To 10 Press SelectClear Output Protection Functions Couple Output ProtectionSelf-Test System-Related OperationsInstrument Identification Instrument State StorageOutput Groups Front Panel Scpi Command Select System\GroupsSystgrodelall SystrebFront Panel Scpi Command Select System\Preferences\Keys Front Panel Scpi Command Select System\Preferences\LockFront Panel Keys LockoutScreen Saver Front Panel DisplayContrast ViewProgramming High-Speed Test Extensions When the power system is turned offList Function All models Refer to , Model DifferencesFront Panel Scpi Command Select Transient\List\Pace Front Panel Scpi Command Select Transient\List\ConfigProgram an Output Pulse or Pulse Train Voltmode LIST, @1Front Panel Scpi Command Select Transient\TrigSource Front Panel Scpi Command Select Transient\List\RepeatTrigtransour BUS, @1 Inittran @1Select the List Step number Program an Arbitrary ListListstep AUTO, @1 Front Panel Scpi Command Select the Transient\List\RepeatFront Panel Scpi Command Select Measure\Sweep Adjust the Measurement Sample RateDigitizer Function Programming the DigitizerTo offset the measurement on Specify a Window FunctionAcquisition trigger, use Front Panel Scpi Command Not Available Front Panel Scpi Command Select Measure\WindowSynchronizing Digitizer Measurements Retrieve Measurement Array DataSensfunc VOLT,@14 Select the Measurement Trigger SourceSensfunc CURR,@14 TransientnInitiate the Measurement Trigger System Front Panel Scpi Command Select Measure\ControlTrigger the Measurement Trigacq @1Appendix a Specifications Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A Performance SpecificationsSupplemental Characteristics Up-programming Settling Time with full resistive load Up-programming Time with full resistive loadDown-programming Settling Time with no load Down-programming Time with Capacitive loadOutput Impedance Graphs Autoranging CharacteristicModel N6753A, CV Mode, @20V, 15A 0125 Magnitude Agilent Models N6731B N6736B and N6741B N6746B Load Transient Recovery TimeFrom 20 Hz 20 MHz from either output to chassis Temperature Coefficient per CCommon Mode Noise Agilent Models N6773A N6776A 20 mA Agilent N6700B, N6701A, N6702A MPS Mainframes Net Weight DimensionsEnvironmental Conditions AC Input432.5 mm 17.03 482.6 mm 19.00 425.45 mm Outline DiagramAppendix B Using the Digital Port Digital Control Port Bi-directional Digital I/O Configuring the Digital Control PortPin Bit Select System\IO\DigPort\PinnExternal Trigger Front Panel Scpi Command Select System\IO\DigPort\PinnDigital Input Front Panel Scpi Command Select System\IO\DigPort\Pin1 Fault OutputFront Panel Scpi Command Select System\IO\DigPort\Pin3 Inhibit InputMainframes Fault/Inhibit System ProtectionClearing a System Protection Fault For the majority of Agilent N6700 Modular Power System This chapter discusses the power allocation functionPower rating of Agilent N6700 mainframes is as follows Rating of the mainframeMainframe Power Limit Power Limit OperationModule Power Limit Must reset the power limit to its maximum rated valuePower limit function will not activate Module Power AllocationTo set a power limit on output Page Appendix D Output On/Off Synchronization Tutorial Synchronizing Output Turn-on DelaysPower Modules Options and Mode Minimum Enable the Synchronization Function ProcedureOutpcoup on Outpcoup OFFSpecify which Output Channels will be Synchronized Specify the Turn-On Delays for each Output ChannelSpecify the Common Delay Offset Synchronizing Multiple Mainframes Digital Connections and ConfigurationSystem\IO\DigPort\Pin6 OperationDIGPIN6FUNC ONC DIGPIN7FUNC OffcIndex Instrument Inhibit input11, 42 ListSystem protection SpecificationsRemote interface Declaration of Conformity Manual Updates