Agilent Technologies N6700 manual Connecting to the Interfaces, Gpib Interface

Page 40

3 Getting Started

Connecting to the Interfaces

 

Electrostatic discharges greater than 1 kV near the interface connectors may

CAUTION

cause the unit to reset and require operator intervention.

 

 

 

 

The Agilent N6700 MPS supports GPIB, LAN, and USB interfaces. All

 

three interfaces are live at power-on. The front panel IO indicator

 

comes on whenever there is activity on the remote interfaces.

GPIB Interface

 

 

 

 

For detailed information about GPIB interface connections, refer to the Agilent

NOTE

 

 

Technologies USB/LAN/GPIB Interfaces Connectivity Guide, located on the

 

 

 

 

 

 

 

 

Automation-Ready CD that is shipped with your product.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The following steps will help you quickly get started connecting your

 

 

 

 

instrument to the General Purpose Interface Bus (GPIB). The

 

 

 

 

following figure illustrates a typical GPIB interface system.

 

 

 

 

 

 

 

 

 

 

 

GPIB Cable

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PC

Instrument

Instrument

 

 

Connect to GPIB Interface

Connect to GPIB

 

Card installed in PC.

port on instrument.

 

1If you have not already done so, install the Agilent IO Libraries Suite from the Automation-Ready CD that is shipped with your product.

2If you do not have a GPIB interface card installed on your computer, turn off your computer and install the GPIB card.

3Connect your instrument to the GPIB interface card using a GPIB interface cable.

4Use the Connection Expert utility of the Agilent IO Libraries Suite to configure the installed GPIB interface card’s parameters.

5The power system is shipped with its GPIB address set to 5. Use the front panel menu if you need to change the GPIB address.

a Press the Menu key, then use the navigation keys to select System\IO\GPIB.

b Use the numeric keys to enter a new value. Valid addresses are from 0 to 30. Press the Enter key to enter the value. Press the Meter key to exit the menu.

6You can now use Interactive IO within the Connection Expert to communicate with your instrument, or you can program your instrument using the various programming environments.

40

Series N6700 User’s Guide

Image 40
Contents Agilent Technologies Low-Profile Modular Power System Legal Notices Agilent Technologies, IncSafety Notices GeneralThis Book Contents Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A 100103 104Essential operating features of the power system Modular Power System MPSReferred to as MPS and power system throughout this manual Agilent N6700 Modular Power System At a Glance Output FeaturesSystem Features Model DifferencesFront Panel At a Glance Rear Panel At a GlanceIndicators Front Panel Display At a GlanceSingle-channel view Interface statusOperating status Front Panel Keys At a GlanceNumber keys Front Panel Menu Reference OCPUSB Subsystem Commands Scpi Command SummaryScpi Command Description Scpi Command FUNCtion VOLTage CURRent, @chanlist SWEepALL Common Commands RST SettingsCommand Description These settings are set by the *RST Reset commandInstallation General Information ModelsItems Supplied OptionsInstalling the Unit Safety ConsiderationsRack Installation Inspecting the UnitInstallation Bench Installation CleaningChannel Number Hz OperationConnecting the Line Cord InstallationPower cord, resulting in fire Snap-On Ferrite CoreConnecting the Outputs Screws securely tightenedTerminals. Sense jumpers are provided for local sensing Terminals in any other wayWire Size AWGResponse Time with an External Capacitor Multiple LoadsPositive and Negative Voltages N6700 Modular Power System Scope Remote Sense Connections Drop in the sense leads can degrade the voltage regulationParallel Connections Open Sense LeadsOver-voltage Protection Considerations Output Noise ConsiderationsWhen programming via the front panel or using Scpi commands Grouping the OutputsFirmware revision B.00.00 and up Discussed in under System-Related OperationsEffect on Specifications Recovery TimeSeries Connections Terminal may be more than 240 VDC from chassis groundSetting the Outputs Desired operating voltageEffect on Specifications Getting Started Menu structure is found in ChapterEntering an Output Voltage Setting Turning the Unit OnSelecting an Output Channel Method 1 Use the Navigation and Arrow KeysEntering a Current Limit Setting Use the On/Off key to enable the outputEnabling the Output Method 2 Use the Current key to enter a valueUsing the Front Panel Menu Press the Menu key to access the front panel command menuSet the Over-Voltage Protection MenuExiting the Command Menu Meter Back There are two ways to exit the command menuCase of Trouble Enter ChannelConnecting to the Interfaces Cause the unit to reset and require operator interventionGpib Interface Automation-Ready CD that is shipped with your productUSB-enabled instrument to the Universal Serial Bus USB. USB InterfaceLAN Interface Following figure illustrates a typical USB interface systemConnecting to a Site LAN Connecting to a Private LANRelease the previous settings Factory-shipped instrument LAN settings are configured toDhcp server. You can leave these settings as they are. Most Will come on when the LAN port has been configuredViewing the Currently Active LAN Settings Configuring the LAN ParametersLAN Parameters Take effectName DomainReset LAN CommunicationUsing the Web Server ReducedUsing Sockets Using TelnetTelnet connections to be made Restoring the Non-volatile Factory Settings Enable/Disable the USB, LAN, and Web ServerSecuring the Interfaces Factory-shipped non-volatile LAN settings Other factory-shipped non-volatile settingsOperating the Power System Programming the Output Select an Output ChannelSet the Output Voltage Set the Voltage Slew RateEnable the Output Set the Output CurrentSequence Multiple Outputs Outpdelrise .1,@2Front Panel Scpi Command Select Output\Polarity Program the Output RelaysOutp Outprelpol REV,@1Enable the Output to Respond to Trigger Commands Front Panel Scpi Command Select Transient\ModeFront Panel Scpi Command Select Transient\Step Synchronizing Output StepsSelect the Transient Trigger Source Initiate the Transient Trigger SystemGenerate Trigger Out Signals Front Panel Scpi Command Select Transient\ControlTrigger the Output Scroll to and select AbortSimultaneous Voltage and Current Measurements Making MeasurementsMeasurement Ranges Using the Protection Functions Front Panel Scpi Command Select Protect\OVPSet the Over-Current Protection Enter a value in the OVP level box To 10 Press SelectCouple Output Protection Clear Output Protection FunctionsSystem-Related Operations Self-TestInstrument Identification Instrument State StorageFront Panel Scpi Command Select System\Groups Output GroupsSystgrodelall SystrebFront Panel Scpi Command Select System\Preferences\Lock Front Panel Scpi Command Select System\Preferences\KeysFront Panel Keys LockoutFront Panel Display Screen SaverContrast ViewWhen the power system is turned off Programming High-Speed Test ExtensionsList Function All models Refer to , Model DifferencesFront Panel Scpi Command Select Transient\List\Config Front Panel Scpi Command Select Transient\List\PaceProgram an Output Pulse or Pulse Train Voltmode LIST, @1Front Panel Scpi Command Select Transient\List\Repeat Front Panel Scpi Command Select Transient\TrigSourceTrigtransour BUS, @1 Inittran @1Program an Arbitrary List Select the List Step numberFront Panel Scpi Command Select the Transient\List\Repeat Liststep AUTO, @1Adjust the Measurement Sample Rate Front Panel Scpi Command Select Measure\SweepDigitizer Function Programming the DigitizerAcquisition trigger, use Specify a Window FunctionTo offset the measurement on Front Panel Scpi Command Select Measure\Window Front Panel Scpi Command Not AvailableSynchronizing Digitizer Measurements Retrieve Measurement Array DataSelect the Measurement Trigger Source Sensfunc VOLT,@14Sensfunc CURR,@14 TransientnFront Panel Scpi Command Select Measure\Control Initiate the Measurement Trigger SystemTrigger the Measurement Trigacq @1Appendix a Specifications Performance Specifications Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762ASupplemental Characteristics Up-programming Time with full resistive load Up-programming Settling Time with full resistive loadDown-programming Settling Time with no load Down-programming Time with Capacitive loadAutoranging Characteristic Output Impedance GraphsModel N6753A, CV Mode, @20V, 15A 0125 Magnitude Load Transient Recovery Time Agilent Models N6731B N6736B and N6741B N6746BCommon Mode Noise Temperature Coefficient per CFrom 20 Hz 20 MHz from either output to chassis Agilent Models N6773A N6776A 20 mA Agilent N6700B, N6701A, N6702A MPS Mainframes Dimensions Net WeightEnvironmental Conditions AC InputOutline Diagram 432.5 mm 17.03 482.6 mm 19.00 425.45 mmAppendix B Using the Digital Port Digital Control Port Configuring the Digital Control Port Bi-directional Digital I/OPin Bit Select System\IO\DigPort\PinnDigital Input Front Panel Scpi Command Select System\IO\DigPort\PinnExternal Trigger Fault Output Front Panel Scpi Command Select System\IO\DigPort\Pin1Front Panel Scpi Command Select System\IO\DigPort\Pin3 Inhibit InputClearing a System Protection Fault Fault/Inhibit System ProtectionMainframes This chapter discusses the power allocation function For the majority of Agilent N6700 Modular Power SystemPower rating of Agilent N6700 mainframes is as follows Rating of the mainframePower Limit Operation Mainframe Power LimitModule Power Limit Must reset the power limit to its maximum rated valueTo set a power limit on output Module Power AllocationPower limit function will not activate Page Appendix D Output On/Off Synchronization Power Modules Options and Mode Minimum Synchronizing Output Turn-on DelaysTutorial Procedure Enable the Synchronization FunctionOutpcoup on Outpcoup OFFSpecify the Common Delay Offset Specify the Turn-On Delays for each Output ChannelSpecify which Output Channels will be Synchronized Digital Connections and Configuration Synchronizing Multiple MainframesOperation System\IO\DigPort\Pin6DIGPIN6FUNC ONC DIGPIN7FUNC OffcIndex Inhibit input Instrument11, 42 ListRemote interface SpecificationsSystem protection Declaration of Conformity Manual Updates