Agilent Technologies N6700 manual Up-programming Time with full resistive load

Page 78

Appendix A Specifications

Supplemental Characteristics (continued)

 

N6751A / N6752A

N6753A

N6754A

N6761A / N6762A

 

 

 

 

Up-programming Time with full resistive load:

 

 

 

(time from 10% to 90% of total voltage excursion)

 

 

 

Small voltage step

0 V to 10 V

0 V to 6 V

0 V to 15 V

0 V to 10 V

Time

0.2 ms

0.4 ms

0.35 ms

0.6 ms

Large voltage step

0 V to 50 V

0 V to 20 V

0 V to 60 V

0 V to 50 V

Time

1.5 ms

1.5 ms

2 ms

2.2 ms

Up-programming Settling Time with full resistive load:

 

 

 

(time from start of voltage change to output voltage within 0.1% of full scale value)

 

 

Small voltage step

0 V to 10 V

0 V to 6 V

0 V to 15 V

0 V to 10 V

Time

0.5 ms

0.8 ms

0.8 ms

0.9 ms

Large voltage step

0 V to 50 V

0 V to 20 V

0 V to 60 V

0 V to 50 V

Time

4 ms

3 ms

4.2 ms

4 ms

Down-programming Time with no load:

 

 

 

 

(time from start of voltage change to output voltage < 0.5 V)

 

 

 

Small voltage step

10 V to 0 V

6 V to 0 V

15 V to 0 V

10 V to 0 V

Time

0.3 ms

0.55 ms

0.6 ms

0.3 ms

Large voltage step

50 V to 0 V

20 V to 0 V

60 V to 0 V

50 V to 0 V

Time

1.3 ms

1.8 ms

2.2 ms

1.3 ms

Down-programming Settling Time with no load:

(time from start of voltage change to output voltage within 0.1% of full scale value)

 

 

Small voltage step

10 V to 0 V

6 V to 0 V

15 V to 0 V

10 V to 0 V

Time

0.45 ms

0.8 ms

0.8 ms

0.45 ms

Large voltage step

50 V to 0 V

20 V to 0 V

60 V to 0 V

50 V to 0 V

Time

1.4 ms

2 ms

2.3 ms

1.4 ms

Down-programming Time with Capacitive load:

(time from start of voltage change to output voltage < 0.5 V)

 

 

 

Small voltage step

10 V to 0 V

6 V to 0 V

15 V to 0 V

10 V to 0 V

Time

2.1 ms

2.2 ms

2.3 ms

4.5 ms

Large voltage step

50 V to 0 V

20 V to 0 V

60 V to 0 V

50 V to 0 V

Time

11 ms

8.5 ms

10 ms

23 ms

Capacitive load

NOTE 2

NOTE 3

NOTE 4

NOTE 2

 

1000 F

4700 F

680 F

1000 F

Down-programming Capability:

 

 

 

 

Continuous power

7 W

12.5 W

12.5 W

7 W

Peak current

7 A

15 A

6 A

3.8 A

High-Speed Test Extensions:

High-speed test extensions are included in Models N676xA and are available as Option 054 in Models N675xA.

Maximum number of List steps = 512 Maximum List dwell time in seconds = 262 Maximum List repetitions = 256, or infinite Maximum measurement points = 4096 Maximum sample rate = 50 kHz

2Modules can discharge a 1000∝F capacitor from full scale to 0V at a rate of 4 times/second.

3Modules can discharge a 4700∝F capacitor from full scale to 0V at a rate of 4 times/second.

4Modules can discharge a 680∝F capacitor from full scale to 0V at a rate of 4 times/second.

78

Series N6700 User’s Guide

Image 78
Contents Agilent Technologies Low-Profile Modular Power System Legal Notices Agilent Technologies, IncSafety Notices GeneralThis Book Contents 103 Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A100 104Modular Power System MPS Essential operating features of the power systemReferred to as MPS and power system throughout this manual Agilent N6700 Modular Power System At a Glance Output FeaturesSystem Features Model DifferencesFront Panel At a Glance Rear Panel At a GlanceSingle-channel view IndicatorsFront Panel Display At a Glance Interface statusFront Panel Keys At a Glance Operating statusNumber keys Front Panel Menu Reference OCPUSB Scpi Command Summary Subsystem CommandsScpi Command Description Scpi Command FUNCtion VOLTage CURRent, @chanlist SWEepALL Command Description Common CommandsRST Settings These settings are set by the *RST Reset commandInstallation Items Supplied General InformationModels OptionsRack Installation Installing the UnitSafety Considerations Inspecting the UnitInstallation Channel Number Bench InstallationCleaning Hz OperationPower cord, resulting in fire Connecting the Line CordInstallation Snap-On Ferrite CoreTerminals. Sense jumpers are provided for local sensing Connecting the OutputsScrews securely tightened Terminals in any other wayWire Size AWGMultiple Loads Response Time with an External CapacitorPositive and Negative Voltages N6700 Modular Power System Scope Remote Sense Connections Drop in the sense leads can degrade the voltage regulationOver-voltage Protection Considerations Parallel ConnectionsOpen Sense Leads Output Noise ConsiderationsFirmware revision B.00.00 and up When programming via the front panel or using Scpi commandsGrouping the Outputs Discussed in under System-Related OperationsSeries Connections Effect on SpecificationsRecovery Time Terminal may be more than 240 VDC from chassis groundSetting the Outputs Desired operating voltageEffect on Specifications Getting Started Menu structure is found in ChapterSelecting an Output Channel Entering an Output Voltage SettingTurning the Unit On Method 1 Use the Navigation and Arrow KeysEnabling the Output Entering a Current Limit SettingUse the On/Off key to enable the output Method 2 Use the Current key to enter a valueSet the Over-Voltage Protection Using the Front Panel MenuPress the Menu key to access the front panel command menu MenuCase of Trouble Exiting the Command MenuMeter Back There are two ways to exit the command menu Enter ChannelGpib Interface Connecting to the InterfacesCause the unit to reset and require operator intervention Automation-Ready CD that is shipped with your productLAN Interface USB-enabled instrument to the Universal Serial Bus USB.USB Interface Following figure illustrates a typical USB interface systemConnecting to a Site LAN Connecting to a Private LANDhcp server. You can leave these settings as they are. Most Release the previous settingsFactory-shipped instrument LAN settings are configured to Will come on when the LAN port has been configuredLAN Parameters Viewing the Currently Active LAN SettingsConfiguring the LAN Parameters Take effectName DomainReset LAN CommunicationUsing the Web Server ReducedUsing Telnet Using SocketsTelnet connections to be made Enable/Disable the USB, LAN, and Web Server Restoring the Non-volatile Factory SettingsSecuring the Interfaces Factory-shipped non-volatile LAN settings Other factory-shipped non-volatile settingsOperating the Power System Set the Output Voltage Programming the OutputSelect an Output Channel Set the Voltage Slew RateSequence Multiple Outputs Enable the OutputSet the Output Current Outpdelrise .1,@2Outp Front Panel Scpi Command Select Output\PolarityProgram the Output Relays Outprelpol REV,@1Front Panel Scpi Command Select Transient\Step Enable the Output to Respond to Trigger CommandsFront Panel Scpi Command Select Transient\Mode Synchronizing Output StepsSelect the Transient Trigger Source Initiate the Transient Trigger SystemTrigger the Output Generate Trigger Out SignalsFront Panel Scpi Command Select Transient\Control Scroll to and select AbortMaking Measurements Simultaneous Voltage and Current MeasurementsMeasurement Ranges Set the Over-Current Protection Using the Protection FunctionsFront Panel Scpi Command Select Protect\OVP Enter a value in the OVP level box To 10 Press SelectCouple Output Protection Clear Output Protection FunctionsInstrument Identification System-Related OperationsSelf-Test Instrument State StorageSystgrodelall Front Panel Scpi Command Select System\GroupsOutput Groups SystrebFront Panel Keys Front Panel Scpi Command Select System\Preferences\LockFront Panel Scpi Command Select System\Preferences\Keys LockoutContrast Front Panel DisplayScreen Saver ViewList Function When the power system is turned offProgramming High-Speed Test Extensions All models Refer to , Model DifferencesProgram an Output Pulse or Pulse Train Front Panel Scpi Command Select Transient\List\ConfigFront Panel Scpi Command Select Transient\List\Pace Voltmode LIST, @1Trigtransour BUS, @1 Front Panel Scpi Command Select Transient\List\RepeatFront Panel Scpi Command Select Transient\TrigSource Inittran @1Program an Arbitrary List Select the List Step numberFront Panel Scpi Command Select the Transient\List\Repeat Liststep AUTO, @1Digitizer Function Adjust the Measurement Sample RateFront Panel Scpi Command Select Measure\Sweep Programming the DigitizerSpecify a Window Function Acquisition trigger, useTo offset the measurement on Synchronizing Digitizer Measurements Front Panel Scpi Command Select Measure\WindowFront Panel Scpi Command Not Available Retrieve Measurement Array DataSensfunc CURR,@14 Select the Measurement Trigger SourceSensfunc VOLT,@14 TransientnTrigger the Measurement Front Panel Scpi Command Select Measure\ControlInitiate the Measurement Trigger System Trigacq @1Appendix a Specifications Performance Specifications Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762ASupplemental Characteristics Down-programming Settling Time with no load Up-programming Time with full resistive loadUp-programming Settling Time with full resistive load Down-programming Time with Capacitive loadAutoranging Characteristic Output Impedance GraphsModel N6753A, CV Mode, @20V, 15A 0125 Magnitude Load Transient Recovery Time Agilent Models N6731B N6736B and N6741B N6746BTemperature Coefficient per C Common Mode NoiseFrom 20 Hz 20 MHz from either output to chassis Agilent Models N6773A N6776A 20 mA Agilent N6700B, N6701A, N6702A MPS Mainframes Environmental Conditions DimensionsNet Weight AC InputOutline Diagram 432.5 mm 17.03 482.6 mm 19.00 425.45 mmAppendix B Using the Digital Port Digital Control Port Pin Bit Configuring the Digital Control PortBi-directional Digital I/O Select System\IO\DigPort\PinnFront Panel Scpi Command Select System\IO\DigPort\Pinn Digital InputExternal Trigger Front Panel Scpi Command Select System\IO\DigPort\Pin3 Fault OutputFront Panel Scpi Command Select System\IO\DigPort\Pin1 Inhibit InputFault/Inhibit System Protection Clearing a System Protection FaultMainframes Power rating of Agilent N6700 mainframes is as follows This chapter discusses the power allocation functionFor the majority of Agilent N6700 Modular Power System Rating of the mainframeModule Power Limit Power Limit OperationMainframe Power Limit Must reset the power limit to its maximum rated valueModule Power Allocation To set a power limit on outputPower limit function will not activate Page Appendix D Output On/Off Synchronization Synchronizing Output Turn-on Delays Power Modules Options and Mode MinimumTutorial Outpcoup on ProcedureEnable the Synchronization Function Outpcoup OFFSpecify the Turn-On Delays for each Output Channel Specify the Common Delay OffsetSpecify which Output Channels will be Synchronized Digital Connections and Configuration Synchronizing Multiple MainframesDIGPIN6FUNC ONC OperationSystem\IO\DigPort\Pin6 DIGPIN7FUNC OffcIndex 11, 42 Inhibit inputInstrument ListSpecifications Remote interfaceSystem protection Declaration of Conformity Manual Updates