Agilent Technologies N6700 manual Usb

Page 14

1 Quick Reference

Menu Command

 

 

 

Control Description

 

 

 

 

 

 

System

IO

USB

Status

 

Displays status, speed, packets received, and packets sent.

 

 

 

Identification

USB connect string - the instrument’s unique USB identifier.

 

 

 

 

 

 

 

 

GPIB

 

 

Selects the GPIB address.

 

 

DigPort

Pin1

Function

Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, FaultOut.

 

 

 

 

Polarity

Specifies the pin polarity: Positive, Negative

 

 

 

Pin2

Function

Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut.

 

 

 

 

Polarity

Specifies the pin polarity: Positive, Negative

 

 

 

Pin3

Function

Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, InhibitIn.

 

 

 

 

Polarity

Specifies the pin polarity: Positive, Negative

 

 

 

Pin4

Function

Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, OnC, OffC.

 

 

 

 

Polarity

Specifies the pin polarity: Positive, Negative

 

 

 

Pin5

Function

Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, OnC, OffC.

 

 

 

 

Polarity

Specifies the pin polarity: Positive, Negative

 

 

 

Pin6

Function

Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, OnC, OffC.

 

 

 

 

Polarity

Specifies the pin polarity: Positive, Negative

 

 

 

Pin7

Function

Specifies the pin function: DigIn, DigIO, TrigIn, TrigOut, OnC, OffC.

 

 

 

 

Polarity

Specifies the pin polarity: Positive, Negative

 

 

 

Data

 

Sends/reads data from the digital I/O port

 

Groups

 

 

 

Defines groups of output channels that are connected in parallel.

 

Preferences

Display

Contrast

 

Configures the display contrast.

 

 

 

 

 

 

 

 

 

Saver

 

Configures the screen saver and wake-on I/O timer.

 

 

 

View

 

Selects 1-channel or 4-channel view at turn-on

 

 

 

 

 

 

 

 

Keys

 

 

Enables/disables key clicks and configures the On/Off key.

 

 

Lock

 

 

Locks front panel keys. Enter a password to unlock the front panel.

 

Admin

Login/Logout

 

Enter a password to access the admin functions.

 

 

 

 

 

 

 

 

Cal

Function

Vprog

High Enters measured data for the High calibration point.

 

 

 

 

 

Low Enters measured data for the Low calibration point.

 

 

 

 

Vmeas

Enters measured data.

 

 

 

 

CMRR

Calibrates common mode rejection ratio.

 

 

 

 

 

 

 

 

 

 

Iprog

High Enters measured data for the High calibration point.

 

 

 

 

 

Low Enters measured data for the Low calibration point.

 

 

 

 

Imeas

100mA Enters measured data for the 100 mA range point.

 

 

 

 

 

100uA Enters measured data for the 100 A range point.

 

 

 

 

Dprog

Calibrates the downprogrammer.

 

 

 

 

Ipeak

Calibrates I peak.

 

 

 

 

 

 

 

 

 

Date

 

Saves the calibration date for each channel.

 

 

 

Save

 

Saves the calibration data.

 

 

LAN

 

 

Enables/disables the LAN interface and the built-in Web server.

 

 

 

 

 

 

 

 

USB

 

 

Enables/disables the USB interface.

 

 

Nvram

 

 

Resets all non-volatile RAM settings to their factory defaults.

 

 

 

 

 

 

 

 

Password

 

 

Changes the password for the admin functions.

 

 

 

 

 

 

 

About

Frame

 

 

Displays model, serial number, and firmware revisions.

 

 

Module

 

 

Displays model, serial number, options, voltage, current, power.

 

 

 

 

 

 

14

Series N6700 User’s Guide

Image 14
Contents Agilent Technologies Low-Profile Modular Power System Legal Notices Agilent Technologies, IncSafety Notices GeneralThis Book Contents 103 Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A100 104Referred to as MPS and power system throughout this manual Modular Power System MPSEssential operating features of the power system Agilent N6700 Modular Power System At a Glance Output FeaturesSystem Features Model DifferencesFront Panel At a Glance Rear Panel At a GlanceSingle-channel view IndicatorsFront Panel Display At a Glance Interface statusNumber keys Front Panel Keys At a GlanceOperating status Front Panel Menu Reference OCPUSB Scpi Command Description Scpi Command SummarySubsystem Commands Scpi Command FUNCtion VOLTage CURRent, @chanlist SWEepALL Command Description Common CommandsRST Settings These settings are set by the *RST Reset commandInstallation Items Supplied General InformationModels OptionsRack Installation Installing the UnitSafety Considerations Inspecting the UnitInstallation Channel Number Bench InstallationCleaning Hz OperationPower cord, resulting in fire Connecting the Line CordInstallation Snap-On Ferrite CoreTerminals. Sense jumpers are provided for local sensing Connecting the OutputsScrews securely tightened Terminals in any other wayWire Size AWGPositive and Negative Voltages Multiple LoadsResponse Time with an External Capacitor N6700 Modular Power System Scope Remote Sense Connections Drop in the sense leads can degrade the voltage regulationOver-voltage Protection Considerations Parallel ConnectionsOpen Sense Leads Output Noise ConsiderationsFirmware revision B.00.00 and up When programming via the front panel or using Scpi commandsGrouping the Outputs Discussed in under System-Related OperationsSeries Connections Effect on SpecificationsRecovery Time Terminal may be more than 240 VDC from chassis groundSetting the Outputs Desired operating voltageEffect on Specifications Getting Started Menu structure is found in ChapterSelecting an Output Channel Entering an Output Voltage SettingTurning the Unit On Method 1 Use the Navigation and Arrow KeysEnabling the Output Entering a Current Limit SettingUse the On/Off key to enable the output Method 2 Use the Current key to enter a valueSet the Over-Voltage Protection Using the Front Panel MenuPress the Menu key to access the front panel command menu MenuCase of Trouble Exiting the Command MenuMeter Back There are two ways to exit the command menu Enter ChannelGpib Interface Connecting to the InterfacesCause the unit to reset and require operator intervention Automation-Ready CD that is shipped with your productLAN Interface USB-enabled instrument to the Universal Serial Bus USB.USB Interface Following figure illustrates a typical USB interface systemConnecting to a Site LAN Connecting to a Private LANDhcp server. You can leave these settings as they are. Most Release the previous settingsFactory-shipped instrument LAN settings are configured to Will come on when the LAN port has been configuredLAN Parameters Viewing the Currently Active LAN SettingsConfiguring the LAN Parameters Take effectName DomainReset LAN CommunicationUsing the Web Server ReducedTelnet connections to be made Using TelnetUsing Sockets Securing the Interfaces Enable/Disable the USB, LAN, and Web ServerRestoring the Non-volatile Factory Settings Factory-shipped non-volatile LAN settings Other factory-shipped non-volatile settingsOperating the Power System Set the Output Voltage Programming the OutputSelect an Output Channel Set the Voltage Slew RateSequence Multiple Outputs Enable the OutputSet the Output Current Outpdelrise .1,@2Outp Front Panel Scpi Command Select Output\PolarityProgram the Output Relays Outprelpol REV,@1Front Panel Scpi Command Select Transient\Step Enable the Output to Respond to Trigger CommandsFront Panel Scpi Command Select Transient\Mode Synchronizing Output StepsSelect the Transient Trigger Source Initiate the Transient Trigger SystemTrigger the Output Generate Trigger Out SignalsFront Panel Scpi Command Select Transient\Control Scroll to and select AbortMeasurement Ranges Making MeasurementsSimultaneous Voltage and Current Measurements Set the Over-Current Protection Using the Protection FunctionsFront Panel Scpi Command Select Protect\OVP Enter a value in the OVP level box To 10 Press SelectCouple Output Protection Clear Output Protection FunctionsInstrument Identification System-Related OperationsSelf-Test Instrument State StorageSystgrodelall Front Panel Scpi Command Select System\GroupsOutput Groups SystrebFront Panel Keys Front Panel Scpi Command Select System\Preferences\LockFront Panel Scpi Command Select System\Preferences\Keys LockoutContrast Front Panel DisplayScreen Saver ViewList Function When the power system is turned offProgramming High-Speed Test Extensions All models Refer to , Model DifferencesProgram an Output Pulse or Pulse Train Front Panel Scpi Command Select Transient\List\ConfigFront Panel Scpi Command Select Transient\List\Pace Voltmode LIST, @1Trigtransour BUS, @1 Front Panel Scpi Command Select Transient\List\RepeatFront Panel Scpi Command Select Transient\TrigSource Inittran @1Program an Arbitrary List Select the List Step numberFront Panel Scpi Command Select the Transient\List\Repeat Liststep AUTO, @1Digitizer Function Adjust the Measurement Sample RateFront Panel Scpi Command Select Measure\Sweep Programming the DigitizerTo offset the measurement on Specify a Window FunctionAcquisition trigger, use Synchronizing Digitizer Measurements Front Panel Scpi Command Select Measure\WindowFront Panel Scpi Command Not Available Retrieve Measurement Array DataSensfunc CURR,@14 Select the Measurement Trigger SourceSensfunc VOLT,@14 TransientnTrigger the Measurement Front Panel Scpi Command Select Measure\ControlInitiate the Measurement Trigger System Trigacq @1Appendix a Specifications Performance Specifications Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762ASupplemental Characteristics Down-programming Settling Time with no load Up-programming Time with full resistive loadUp-programming Settling Time with full resistive load Down-programming Time with Capacitive loadAutoranging Characteristic Output Impedance GraphsModel N6753A, CV Mode, @20V, 15A 0125 Magnitude Load Transient Recovery Time Agilent Models N6731B N6736B and N6741B N6746BFrom 20 Hz 20 MHz from either output to chassis Temperature Coefficient per CCommon Mode Noise Agilent Models N6773A N6776A 20 mA Agilent N6700B, N6701A, N6702A MPS Mainframes Environmental Conditions DimensionsNet Weight AC InputOutline Diagram 432.5 mm 17.03 482.6 mm 19.00 425.45 mmAppendix B Using the Digital Port Digital Control Port Pin Bit Configuring the Digital Control PortBi-directional Digital I/O Select System\IO\DigPort\PinnExternal Trigger Front Panel Scpi Command Select System\IO\DigPort\PinnDigital Input Front Panel Scpi Command Select System\IO\DigPort\Pin3 Fault OutputFront Panel Scpi Command Select System\IO\DigPort\Pin1 Inhibit InputMainframes Fault/Inhibit System ProtectionClearing a System Protection Fault Power rating of Agilent N6700 mainframes is as follows This chapter discusses the power allocation functionFor the majority of Agilent N6700 Modular Power System Rating of the mainframeModule Power Limit Power Limit OperationMainframe Power Limit Must reset the power limit to its maximum rated valuePower limit function will not activate Module Power AllocationTo set a power limit on output Page Appendix D Output On/Off Synchronization Tutorial Synchronizing Output Turn-on DelaysPower Modules Options and Mode Minimum Outpcoup on ProcedureEnable the Synchronization Function Outpcoup OFFSpecify which Output Channels will be Synchronized Specify the Turn-On Delays for each Output ChannelSpecify the Common Delay Offset Digital Connections and Configuration Synchronizing Multiple MainframesDIGPIN6FUNC ONC OperationSystem\IO\DigPort\Pin6 DIGPIN7FUNC OffcIndex 11, 42 Inhibit inputInstrument ListSystem protection SpecificationsRemote interface Declaration of Conformity Manual Updates