On the Calibration screen under Step 4 (See Fig. 4.1), there are six parameters that describe the lens that will be analyzed:
1.Diameter
2.Index of Refraction
3.Power
4.Front Curve
5.Back Curve
6.Center Thickness
These parameters are for both the plus and the minus lenses.
NOTE
If these lenses are misplaced or compromised in any way, please call the factory and order a new calibration lens. Once the new lens arrives, check these parameters and make any necessary changes.
To perform this calibration, select the Calibrate button. You will be prompted to remove the lens. Make sure there is no lens in the chamber and select OK. After 25 seconds, you will be prompted to insert the calibration lens. Put the
Select OK.
After both lenses are analyzed, the plus and minus slope adjustments will be displayed. Select YES to accept these changes. The calibration wizard will appear telling you that the power calibration is complete. Select OK. To exit back to the Job screen, select OK at the bottom of the screen and then on the next screen select Next.
NOTE
On this screen, there is also a Power Offset box. This can be used if you see a consistent power error. For example, powers are showing 0.25 weak across the board. Enter a +.25 and all analyzed powers will have 0.25 added to the final reading. This functions like the manual eyepiece adjustment for power on a conventional lensometer.
This screen includes two parameters labeled Minus Prism Bias and Plus Prism Bias. They are constants and should not be changed unless directed by the factory.
To verify power calibration, there is an internal job (Job Number 999999999) which will allow you to use the two calibration lenses (+6 and
Clarity User Guide, March 1, 2011 - Rev. 1.06 | Page 33 |