Clarity 87392-11 manual Teach Screen

Page 18

Figure 2.4 – The Teach Screen

At the lower left of the screen, there may appear one or more images that the Clarity has identified as candidates for the fitting cross. If one of these matches the actual fitting cross and is on axis, left-click the image; it will move to the Image Box at the left top of the screen.

NOTE

Do not select a candidate image from the lower-left unless if is aligned. In Figure 2.4, the image is not aligned so should not have been copied to the Image Box at upper left.

If there is no suitable candidate image at the lower-left, two alternatives remain for creating an image:

1.Use the image cleanup tools and the mouse to draw an image in the image box. The width of drawn pixels may be selected by clicking on one of the tool-width squares. Select the black draw color to actually draw, and the white draw color to clear or erase.

2.Click the Feature From Camera control. A screen shot of the lens will appear. Rotate the lens if necessary to align it. Then use the mouse to draw a rectangle around the fitting cross. The pixels enclosed in the rectangle will be copied to the Image Box. The rectangle should be as small as possible but still include the complete image of the fitting cross.

Once an image has been acquired (from among the candidates, or drawn, or copied from the camera image) it may need to be “cleaned up” by removing stray pixels or adding pixels to fill gaps. Use the Image Cleanup tools for this purpose. (Cleanup is especially important for camera captures.) The result should be a clean image that nearly spans the width and height of the Image Box.

Clarity User Guide, March 1, 2011 - Rev. 1.06

Page 18

Image 18
Contents Clarity User Guide Safety Reminders Table of Contents Home Burn Distance, Near and Prism Power Checks Getting Started Unpacking the ClarityFoam Shipping Blocks in the Blocking Chamber 1a Block #1 P/N 33593 Bottom Shipping Block1c Shipping Bracket Assembly, Clarity P/N 1d Installed Shipping Bracket Assembly Clarity Front View User Items Clarity Rear Connector Bulkhead Installing the Glass Plate This page intentionally left blank for duplexing purposes Powering on the Clarity Basic OperationEstablishing Communications Touch Screen Display and Job Screen Keep hands away from unit during the blocking functionProcessing Screen Processing ScreenAnalysis Screen Analysis ScreenClarity User Guide, March 1, 2011 Rev Teach Screen Clarity User Guide, March 1, 2011 Rev This page intentionally left blank for duplexing purposes Setup Screens Main Menu ScreenPreferences Screen Key Features of the Preferences Screen Preferences ScreenTolerances Screen Tolerances ScreenInterface Screen Interface ScreenBlocking Screen Calibration Screen Calibration ScreenKey Features of the Advanced Screen Advanced ScreenSet Date and Time Technician Screen View Log Copy Logs to Flash DriveStatistics Screen Statistics ScreenThis page left blank intentionally for duplexing purposes Calibration Clarity User Guide, March 1, 2011 Rev Clarity User Guide, March 1, 2011 Rev Touch Screen Calibration Touch Screen Calibration ScreenClarity User Guide, March 1, 2011 Rev This page intentionally left blank for duplexing purposes Advanced Screen in the Setup Menu Distance PowerLens Feature Detection Near PowerPrism Power Warm-Up TimeSetup Password Job Data LogHome This screen intentionally left blank for duplexing purposes Cleaning the LCD MaintenanceMaintenance Tips Cleaning the Touch ScreenThis page left blank intentionally for duplexing purposes Intended USE & Limitations Clarity User Guide, March 1, 2011 Rev Glossary of Terms Clarity User Guide, March 1, 2011 Rev Appendix a General Specifications External ConnectionsTechnical Specifications Appendix B Certifications & StandardsThis page left blank intentionally for duplexing purposes Appendix C Electrical Diagram To be suppliedThis screen intentionally left blank for duplexing purposes Appendix D Certificate of Compliance