Setup Password:
The Setup Password is set at the factory to 082379. This feature allows you to change the password if you choose. If you forget the password, call Technical Support.
Job Data Log:
Activate Job Data Log – Place a check in this box if you would like to begin maintaining a job log.
View Log – Select this button if you would like to view the log. (The log must have been previously activated.) This can be useful in diagnosing issues.
View Blocked Job Log – Select this button if you want to view only those jobs that were successfully blocked.
There are six buttons in the upper center of this screen (See Fig. 5.1):
1.Update Software - Select this button to download new software versions. Follow the
2.Set Date/Time – Use this function to set the time if you are not in the eastern U.S.A. time zone. The date and U.S. Eastern Time are entered at the factory.
3.Restore – If you have made a backup of the system data, you can restore the data if need be.
4.Backup – Follow the
5.Import Lens Data – This function can be used when periodic updates are available to update and import new lens data. This is finished lens data that the Clarity uses to verify power.
6.Save Tech Data – This function causes all of the Setup screens to be saved. Saved screen images may be useful for error recovery.
NOTE
We recommend backing up your system prior to and immediately after making any significant changes in the Setup Menu. The USB thumb drive needs to be 256 megabyte or higher.
Burn In:
In the lower left corner of the screen there is a Burn In button. This button is used to exercise a cold machine or to diagnose axis movement issues. A box will appear with a default number of cycles set for operation. You can change this number or accept it and select OK. Select the Stop button to abort the burn in cycles.
Home:
This button causes the axes to move to their home positions. The Home function is also available from the Technician screen.
Clarity User Guide, March 1, 2011 - Rev. 1.06 | Page 39 |