Troubleshooting - 3
From Sheet 2
Program current to full scale, voltage to Vmax and load to the power supply's rated current. Supply should be in CC.
Will not go into CC or error > 10%
?
No
Output out of spec but close?
No
Output OK but meter wrong?
No
| CC_Prog OK ? |
|
|
| Replace A1 |
|
| Yes |
| ||
Yes | (see Table |
|
| ||
| |||||
|
|
|
|
| |
|
|
|
|
| |
|
|
|
|
|
|
No
Replace A2
|
|
| Calibrate unit |
Yes |
|
| |
| |||
|
|
|
|
|
| Calibrate current. If still |
|
| wrong or will not |
Yes | ||
|
| calibrate, replace A2 |
|
|
|
Turn on OCP and insure Protect trips.
Prot trips? |
|
|
| CC_detect* low? |
|
|
|
| Check cable W9, | ||||||
| No |
| No |
| |||||||||||
|
| (see Table |
|
| replace A1 | ||||||||||
|
|
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
| Yes |
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Replace A2 |
|
|
|
| ||||
| Yes |
|
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Goto Sheet 4
Figure 3-1 Sheet 3. Troubleshooting Flowchart