Agilent Technologies 5962-8200 service manual Sheet 2. Troubleshooting Flowchart

Page 30

3 - Troubleshooting

From Sheet 1

Enable output and program voltage and current full scale with no load. Measure output voltage.

 

 

 

 

 

Check to insure OV

Unit OV's?

 

 

 

setting is not less than

 

Yes

 

 

 

the voltage setting. If not

 

 

 

 

 

 

 

 

then replace A1.

 

 

 

 

 

 

 

 

 

 

 

 

No

Output voltage

 

 

 

Yes

> 10% error?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

CV_Prog &

 

 

 

Check cable W9,

CC_Prog OK? (see

 

 

 

 

Yes

 

 

 

Replace A1

Table 3-4)

 

 

 

 

 

 

 

No

Replace A2

Output out of spec

 

 

 

Calibrate voltage

 

 

Yes

 

 

 

but close?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output OK but

 

 

 

Calibrate voltage. If

 

 

Yes

 

still wrong or will not

 

meter wrong?

 

 

 

 

 

 

calibrate, replace A2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

 

Program the OV 2 volts lower than the output voltage.

Unit OV's? No Program OV to full scale

Yes

OV_Prog OK?

 

 

 

Calibrate OV. If OV is still

 

Yes

 

not functioning properly

(see Table 3-4)

 

 

 

 

 

check W9, replace A1.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No

 

 

 

 

 

Go to Sheet 3

Replace A2

Figure 3-1 Sheet 2. Troubleshooting Flowchart

30

Image 30
Contents Agilent Part No Microfiche No September Warranty Information CertificationSafety Summary Before Applying PowerSafety Symbol Definitions Symbol DescriptionPrinting History Instrument IdentificationTable of Contents Post-repair Calibration Initialization ROM UpgradeDisassembly Procedures IntroductionPage Related Documents Safety ConsiderationsOrganization ChapterManual Revisions RevisionsElectrostatic Discharge Firmware RevisionsIntroduction Test Equipment RequiredTest Setup Type Specifications Recommended ModelMeasurement Techniques Test Setup Electronic LoadCurrent-Monitoring Resistor Operation Verification TestsPerformance Tests ProgrammingVoltage Programming and Readback Accuracy CV SetupConstant Voltage CV Tests CV Load EffectTransient Recovery Time CV Source EffectCV Noise Pard CC Setup Constant Current CC TestsCurrent Programming and Readback Accuracy Current Sink -CC Operation Low Range Current Readback AccuracyCC Load and Line Regulation CC Load EffectCC Source Effect CC Noise PardPerformance Test Equipment Form Constant Current Tests Current Programming and Readback Load Effect Source Effect Pard Ripple and NoiseTransient Response Current Sink @ -3A Readback MA Range Current ReadbackSource Effect Vout − Vout −Load Effect Vout − Vout + 2.0mV Iout −Iout − Iout + 2.2 mA Vout + 4 mVIout − Iout + 0.2 mA Current Sink @ -0.6A Readback Iout −Iout − Iout + 1.1 mA Load Effect Vout − Vout + 5 mVIout − Iout + 0.1 mA Current Sink @ -0.3A Readback Iout −Page Troubleshooting Test Equipment Required for Troubleshooting Overall TroubleshootingFlow Charts Type Purpose Recommended ModelSheet 1. Troubleshooting Flowchart Sheet 2. Troubleshooting Flowchart Sheet 3. Troubleshooting Flowchart Sheet 4. Troubleshooting Flowchart Self-Test Error Codes/Messages Specific Troubleshooting ProceduresPower-on Self-test Failures Error Code Description Probable CauseBias and Reference Voltages CV/CC Status Annunciators TroubleshootingBias and Reference Supplies Bias Test Point MeasurementJ307 Voltage Measurements Voltage Measurements at J207 A2 Interface to A1 Main boardDisabling Protection Features Manual Fan Speed ControlInhibit Calibration Switch Calibration PasswordPost-repair Calibration Initialization ROM UpgradeUpgrade Procedure Identifying the FirmwareDisassembly Procedures List of Required ToolsCover, Removal and Replacement A2 Interface Board, Removal and ReplacementFront Panel Assembly, Removal and Replacement A3 Front Panel Board, Removal and Replacement T1 Power Transformer, Removal and ReplacementA1 Main Control Board Line Voltage Wiring Interface Signals Power Supply Interface signalsConnector Signal Description Primary Interface A3 Front Panel CircuitsA2 Interface Circuits Secondary InterfacePower Circuits A1 Main Board CircuitsControl Circuits Principles of Operation Page Chassis, Electrical Designator PartNumber Qty DescriptionReplaceable Parts List Mechanical Parts Identification Chassis, Mechanical MP9Diagrams Diagrams A2/A3 Boards Block Diagram Bias CircuitsIndex Index