HP 6621A, 6623A, 6624A manual General Information, Introduction, Safety Considerations, Options

Page 11

1

General Information

Introduction

This chapter contains a general description of your power supply, as well as its performance specifications. Information about options, accessories, and HP-IB cables is also provided. This manual describes all five models in the HP 6621A- 6624A, and 6627A power supply family. Unless stated otherwise, the information in this manual applies to all of these models. Information that is specific to one model only is identified as such in this manual.

Safety Considerations

This product is a Safety Class 1 instrument, which means that it is provided with a protective earth terminal. This terminal must be connected to a power source that has a 3-wire ground receptacle. Review the instrument and this manual for safety markings and instructions before operation. Refer to the Safety Summary page at the beginning of this manual for a summary of general safety information. Safety information for specific procedures is located at appropriate places in this manual.

Instrument and Manual Identification

Hewlett-Packard power supplies are identified by a two-part serial number, i.e. 2601A-00101. The first part of the serial number (the prefix) is a number/letter combination that denotes either the date of manufacture or the date of a significant design change. It also indicates the country of origin. (Starting at 1960, 26 = 1986; 01 = the first week of the year; A = U.S.A.) The second part of the serial number is a different sequential number assigned to each instrument starting with 00101.

If the serial number prefix on your power supply differs from that shown on the title page of this manual, a yellow Manual Changes sheet that is supplied with this manual explains the difference between your instrument and the instrument described by this manual. The change sheet can also contain information for correcting errors in the manual.

Options

Options 100,120, 220, and 240 simply determine which line voltage is selected at the factory. For information about changing the line voltage setting, see Line Voltage Conversion, page 28.

Option 750 consists of a fault indicator (FLT) and remote inhibit (INH) circuit and relay control, which provide additional shutdown protection should either the HP-IB and/or controller fail. This Option is described in a separate document entitled, "Appendix E Option 750 Operating Instructions for the Multiple Output Linear System DC Power Supply, HP Models 6621A, 6622A, 6623A, 6624A, and 6627A (HP P/N 5957-6372).

#100 Input power, 100 Vac, 47--66 Hz #120 Input power, 120 Vac, 47--66 Hz #220 Input power, 220 Vac, 47--66 Hz #240 Input power, 240 Vac, 47--66 Hz

#700 Computer Interface Intermediate Language (CIIL) #750 Fault (FLT) Remote Inhibit (INH) and Relay Control #908 One rack mount kit (5062-3977)

#909 One rack mount kit with handles (5062-3983) #910 One service manual with extra operating manual

General Information 11

Image 11
Contents HP Part No Operating ManualCertification Safety Summary Symbol Description Safety SummaryEMC Declaration of ConformityAppendix D--Error Messages Installation ProceduresAppendix A--Calibration Procedure Appendix C--Command SummaryTable Of Contents Local Operation Remote OperationProgramming With The Series 200/300 Computer Command SummaryError Messages CalibrationPage Instrument and Manual Identification Safety ConsiderationsGeneral Information IntroductionOutput Combinations Available Model AccessoriesDescription HP-IB Board Basic OperationOutput Low Range Values High Range Values Output Boards Definitions SpecificationsQualifying Conditions Output Response Characteristics Source Effect SpecificationsOutputs Low High Voltage Temperature Coefficient Supplemental CharacteristicsCommand Processing Time see Figure Outputs Low High Voltage Programming ResolutionReadback Resolution AC Input Power and CurrentLow Voltage High Voltage 80 W Low Voltage Safety Agency ComplianceDimensions all models Output ImpedanceGeneral Information General Information General Information General Information Location and Cooling InstallationInitial Inspection Line Fuse Input Power Requirements100/120 2110-0342 220/240 2110-0055 Line FusesLine Voltage Conversion Power CordHP-IB Interface Connector Page Turning On Your Supply Front Panel Controls and IndicatorsGetting Started HP-IB Status Annunciators LCL keySystem Control Keys Power Supply StatusAnnunciators Alphanumeric LCDOutput Control Keys Normal Self Test Indications Line SwitchNumeric Entry Keys Sample Self-Test Failure Display Checking Out Your Supply Using Local ControlCurrent Test Voltage TestOvervoltage Test Enter/Output Statements Introduction To Remote OperationIset Enter OCPAddr Sending a Remote CommandOutput Reading the HP-IB AddressDisp a Often Used CommandsGetting Data From The Supply Disp a Returning the Supply to Local Mode Output Ranges Output Connections and Operating InformationRange Selection Protection FeaturesOperating Quadrants Typical Output Range Characteristics Connecting the Load Page Cross Section Meters Area in mm2 Wire Bundled 10 a 20 aFeet Positive and Negative Voltages Remote Voltage SensingMultiple Loads Remote Sense Connections Remote Voltage SensingOpen Sense Leads Output Type FormulaOutput Noise Considerations Programming Response Time with an Output CapacitorExternal Trigger Circuit Overvoltage Trigger ConnectionsEquivalent Internal OV Trigger Circuit Parallel Operation Power Supply Protection ConsiderationsBattery Charging CV Operation Maximum Allowable Voltage SettingRemote Sensing CC Operation13. Series Connections with Local Sensing CV Operation Series Operation14. Series Connections with Remote Sensing Specifications for Series OperationPage Page Interface Function Remote OperationHP-IB Operation HP-IB Address Selection Numeric Data Power-On Service Request PONProgramming Syntax Sheet 1 of 2. Syntax Forms for Power Supply Commands Sheet 2 of 2. Syntax Forms for Power Supply Commands Ovset Fault ?Vset IsetPage Initial Conditions Power Supply CommandsCurrent Programming Voltage ProgrammingRange Switching OVSET? Output On/OffOvervoltage OV Protection Multiple Output Storage & Recall Overcurrent Protection OCPStatus Reporting Clear CommandASTS? UNR +CCBit Assignment of the Serial Poll Register FAULT?UNMASK? Unmask 2,XXXService Request Generation PON ? SRQ?Display On/Off Reprogramming DelayRQS Bit Other Queries Response Code Front PanelExplanation Code Explanation TEST? ResponsesPage General Local ModeLocal Operation Local Control Of Output FunctionsSetting Current Setting VoltageDisplaying the Contents of the Fault Register Setting Overvoltage ProtectionResetting Overvoltage Protection Resetting Overcurrent ProtectionCondition Setting the Reprogramming DelaySetting the Supplys HP-IB Address Local Control Of System FunctionsRCL Enter Displaying Error MessagesAddr Enter STO EnterPage Test Equipment and Setup Required Calibration ProceduresFigure A-1. Calibration Setup General Calibration Procedure Are not stored. Exercise care when moving the leads Pause Calibration Program10 ! Calibration Example Clear Voltmeter Output BufferFnend Input ANY More Outputs to CALIBRATE? Y or N,X$Disp END of Calibration Program Page Page Voltage and Current Programming Programming With a Series 200/300 ComputerPath Names Voltage and Current Readback Voltage and Current Programming With VariablesPresent Status Programming Power Supply RegistersPrint OUTPUT1 is in CV Mode END if Service Request and Serial PollPrint Overvoltage on Output #2 Enable IntrOFF Intr Print Overvoltage on Output #1Error Detection Stored Operating States Programming Outputs Connected In ParallelInput Enter Operating VOLTAGE,V1 Input Enter Voltage LIMIT,VInput Enter the Desired Current Limit POINT,I Programming Outputs Connected In SeriesCommand Description Command SummaryTable C-1. Command Summary Table C-l. Command Summary ROM? PON?Previously Test Responses Error Codes and MessagesPower-On Self Test Messages Error Responses Table D-l. Power-On Self Test Error MessageError Code Message Explanation ERR? query ERR key Table D-2. Error ResponsesResponse Code Explanation TEST? query Table D-3. TEST? ResponsesPage Make Changes Manual Backdating6623A Generally Applicable AnnotationsII. CE92 Product Specific Annotations 6621ALatin America Or contactEurope Far East

6624A, 6621A, 6623A specifications

The HP 6623A, 6621A, and 6624A are precision DC power supplies widely recognized for their reliability and performance in various laboratory and industrial applications. These models are part of HP's 662X series, designed to cater to the needs of engineers, researchers, and technicians who require accurate power sources for their testing and development activities.

The main features of the HP 6623A, 6621A, and 6624A include their high stability, low ripple, and excellent load regulation, ensuring that the output voltage and current remain stable during testing. The power supplies provide multiple output channels, allowing users to power multiple devices simultaneously. The flexibility in setting voltage and current levels makes these models ideal for a wide range of applications, including semiconductor testing, device characterization, and system integration.

One of the standout technologies in the HP 662X series is the use of smart design techniques that minimize noise and enhance output performance. These power supplies incorporate advanced feedback mechanisms to maintain steady output, even under varying load conditions. Additionally, they feature programmable outputs, which means users can adjust the output levels through a connected computer or control system, streamlining the testing process and improving efficiency.

The HP 6623A model offers three independent outputs, with total power capabilities of 40 watts. It includes a 0-20V output, which can deliver up to 2A of current, along with two additional outputs that are adjustable. The 6621A provides a single output option, delivering a maximum of 20V and 2A, making it well-suited for simple applications where a single power source is required. In contrast, the HP 6624A stands out with its four independent output channels, providing a total of 60 watts, making it the most versatile of the three models.

Characteristics of these power supplies include user-friendly interfaces, allowing for easy configuration and monitoring of settings. LED indicators provide real-time feedback on voltage and current levels, enabling users to quickly assess the performance of their tests. Moreover, built-in protection features safeguard both the power supply and the connected devices from overvoltage and overcurrent conditions.

Overall, the HP 6623A, 6621A, and 6624A power supplies embody advanced engineering and design, making them invaluable tools for professionals looking for high-quality, reliable power sources for their electronic testing needs.