Table
Command | Header | Channel* | Data | Syntax |
|
| Range |
| (see Figure |
Calibration Mode | CMODE |
| 0,1 (off,on) | C2 |
Set High Voltage | VHI | 1,2,3,4 | - | C3 |
Set Low Voltage | VLO | 1,2,3,4 | - | C3 |
Set High Current | IHI | 1,2,3,4 | - | C3 |
Set Low Current | ILO | 1,2,3,4 | - | C3 |
Calibrate Overvoltage | OVCAL | 1,2,3,4 | - | C3 |
Voltage Data | VDATA | 1,2,3,4 | see Table | C5 |
Current Data | IDATA | 1,2,3,4 | see Table | C5 |
*Channels 3 and 4 are not used in all models.
Table
| Voltage Range in Volts | Current Range in Amps | ||||||
|
| Vlo |
| Vhi | llo |
|
| Ihi |
Output | Min | Max | Min | Max | Min | Max | Min | Max |
40 W Low Voltage | 0 | 0.1 | 18.5 | 20.5 | 0 | 0.15 | 4.5 | 5.3 |
40 W High Voltage | 0.04 | 0.20 | 44 | 48 | 0 | 0.1 | 1.5 | 2.1 |
80 W Low Voltage | 0 | 0.1 | 18.5 | 20.5 | 0.05 | 0.25 | 9 | 10.5 |
80 W High Voltage | 0.04 | 0.20 | 44 | 48 | 0 | 0.1 | 3.5 | 4.2 |
General Calibration Procedure
This procedure causes the voltage of the specified output to go to full scale value. Take appropriate precautions.
The following general procedure applies to any computer that you are using to control your power supply. This procedure must be repeated for each output on your power supply. The calibration commands shown are the actual string commands that must be sent to the power supply. Because the power supply will attempt to calibrate itself even when incorrect or invalid readings are sent to it, it is recommended that you perform some type of error checking after steps 4, 5, and 8 to ensure that the values sent to the supply are legitimate. Use the ranges in Table
Before you continue with this procedure, disconnect all loads from the supply, strap the supply for local sensing, and connect the voltmeter to the + S and - S terminals as shown in the voltage calibration setup of Figure
Start with output channel 1 and use the following commands to calibrate your power supply:
Calibration Procedures 91