HP 6623A Vset, Iset, Ovset, Out, Dcpon, Unmask, Dly, Ovrst, Ocrst, Srq, Pon, Dsp, Sto, Rcl, Clr

Page 66

Table 5-1. Power Supply Commands

Command

Header

*Output Channel

Data Range

Syntax (Fig. 5-2)

Set Voltage

VSET

1,2,3,4

See Table 5-4

C4

Set Current

ISET

1,2,3,4

See Table 5-4

C4

Set Overvoltage

OVSET

1,2,3,4

See Table 5-4

C4

OC Protection On/Off

OCP

1,2,3,4

0,1(off,on)

C4

Output On/Off

OUT

1,2,3,4

0,1(off,on)

C4

Set the State of all

DCPON

----

0,1(off,on) CC+

C4

Outputs at Power-On

 

 

2,3 (off,on) CC-

 

Unmask

UNMASK

1,2,3,4

0-255

C4

Reprogram Delay

DLY

1,2,3,4

0-32 (LSB=0.004)

C4

(in seconds)

 

 

 

 

Reset Overvoltage

OVRST

1,2,3,4

--

C3

Reset Overcurrent

OCRST

1,2,3,4

--

C3

Service Request

SRQ

--

0,1,2,3

C2

Power-On SRQ

PON

--

0,1(off, on)

C2

On/Off

 

 

 

 

Display On/Off

DSP

--

0,1(off, on)

C2

Display Characters

DSP

--

"string"

C6

(up to 12 characters)

 

 

 

 

Store Settings

STO

--

1-10

C2

Recall Settings

RCL

--

1-10

C2

Clear Supply

CLR

--

--

C1

*Output channels 3 and 4 are not used in all models (see Table 5-4).

Table 5-2. Power Supply Queries

Query

Header

Channel

Response

 

Initial Value

Syntax

 

(Note 7)

(Note 1)

(Notes 5 and 6)

 

(Fig. 5-2)

Voltage Setting

VSET ?

1,2,3,4

SZD.DDD

 

0

Q2

Current Setting

ISET ?

1,2,3,4

SZD.DDD

 

Min. Value

Q2

 

 

 

SZZD.DD(Note 2)

(see Table 5-4)

 

Voltage Output

VOUT ?

1,2,3,4

SZD.DDD

 

--

Q2

Current Output

IOUT ?

1,2,3,4

SZD.DDD SD.DDDD

--

Q2

 

 

 

(Note 3)

 

 

 

OVP Setting

OVSET ?

1,2,3,4

SZZD.DD

 

Full Scale

Q2

 

 

 

 

 

(See Table 5-4)

 

OC Protection On/Off

OCP ?

1,2,3,4

ZZD

 

0(0ff)

Q2

Output On/Off

OUT ?

1,2,3,4

ZZD

 

1(0n)

Q2

Unmask Setting

UNMASK ?

1,2,3,4

ZZD

 

0(cleared)

Q2

Delay Setting

DLY ?

1,2,3,4

<sp>ZD.DDD

0.02 sec

Q2

Status

STS ?

1,2,3,4

ZZD

 

--

Q2

Accumulated Status

ASTS ?

1,2,3,4

ZZD

 

--

Q2

Fault

FAULT ?

1,2,3,4

ZZD

 

--

Q2

Error

ERR ?

--

ZZD

 

--

Q1

Service Request Setting

SRQ ?

--

ZZD

 

0(0ff)

Q1

Power-On SRQ On/Off

PON ?

--

ZZD

 

Last stored value

Q1

Display On/Off

DSP ?

--

ZZD

 

1(0n)

Q1

Model Number

ID ?

--

HP 662XA (Note 4)

--

Q1

Selftest

TEST ?

--

ZZD

 

--

Q1

Calibration Mode

CMODE ?

--

ZZD

 

0(0ff)

Q1

S = Sign Z = Digit with leading zeros put out as spaces.

D= Digit

<sp > = space

 

66 Remote Operation

Image 66
Contents Operating Manual HP Part NoCertification Safety Summary Safety Summary Symbol DescriptionDeclaration of Conformity EMCAppendix C--Command Summary Installation ProceduresAppendix A--Calibration Procedure Appendix D--Error MessagesTable Of Contents Remote Operation Local OperationCalibration Command SummaryError Messages Programming With The Series 200/300 ComputerPage Introduction Safety ConsiderationsGeneral Information Instrument and Manual IdentificationAccessories DescriptionOutput Combinations Available Model Basic Operation Output Low Range Values High Range ValuesHP-IB Board Output Boards Specifications Qualifying ConditionsDefinitions Output Response Characteristics Specifications Source EffectSupplemental Characteristics Outputs Low High Voltage Temperature CoefficientAC Input Power and Current Outputs Low High Voltage Programming ResolutionReadback Resolution Command Processing Time see FigureOutput Impedance Safety Agency ComplianceDimensions all models Low Voltage High Voltage 80 W Low VoltageGeneral Information General Information General Information General Information Installation Initial InspectionLocation and Cooling Input Power Requirements Line FuseLine Fuses 100/120 2110-0342 220/240 2110-0055Power Cord Line Voltage ConversionHP-IB Interface Connector Page Front Panel Controls and Indicators Getting StartedTurning On Your Supply LCL key HP-IB Status AnnunciatorsAlphanumeric LCD Power Supply StatusAnnunciators System Control KeysOutput Control Keys Line Switch Numeric Entry KeysNormal Self Test Indications Checking Out Your Supply Using Local Control Sample Self-Test Failure DisplayVoltage Test Overvoltage TestCurrent Test OCP Introduction To Remote OperationIset Enter Enter/Output StatementsReading the HP-IB Address Sending a Remote CommandOutput AddrOften Used Commands Getting Data From The SupplyDisp a Disp a Returning the Supply to Local Mode Output Connections and Operating Information Output RangesProtection Features Operating QuadrantsRange Selection Typical Output Range Characteristics Connecting the Load Page Wire Bundled 10 a 20 a FeetCross Section Meters Area in mm2 Remote Voltage Sensing Multiple LoadsPositive and Negative Voltages Remote Voltage Sensing Remote Sense ConnectionsProgramming Response Time with an Output Capacitor Output Type FormulaOutput Noise Considerations Open Sense LeadsOvervoltage Trigger Connections External Trigger CircuitEquivalent Internal OV Trigger Circuit Power Supply Protection Considerations Battery ChargingParallel Operation Maximum Allowable Voltage Setting CV OperationCC Operation Remote SensingSeries Operation 13. Series Connections with Local Sensing CV OperationSpecifications for Series Operation 14. Series Connections with Remote SensingPage Page Remote Operation HP-IB OperationInterface Function HP-IB Address Selection Power-On Service Request PON Programming SyntaxNumeric Data Sheet 1 of 2. Syntax Forms for Power Supply Commands Sheet 2 of 2. Syntax Forms for Power Supply Commands Iset Fault ?Vset OvsetPage Power Supply Commands Initial ConditionsVoltage Programming Current ProgrammingRange Switching Output On/Off Overvoltage OV ProtectionOVSET? Overcurrent Protection OCP Multiple Output Storage & RecallClear Command Status ReportingUNR +CC ASTS?Unmask 2,XXX FAULT?UNMASK? Bit Assignment of the Serial Poll RegisterService Request Generation SRQ? PON ?Reprogramming Delay RQS BitDisplay On/Off Other Queries Front Panel ExplanationResponse Code TEST? Responses Code ExplanationPage Local Control Of Output Functions Local ModeLocal Operation GeneralSetting Voltage Setting CurrentResetting Overcurrent Protection Setting Overvoltage ProtectionResetting Overvoltage Protection Displaying the Contents of the Fault RegisterLocal Control Of System Functions Setting the Reprogramming DelaySetting the Supplys HP-IB Address ConditionSTO Enter Displaying Error MessagesAddr Enter RCL EnterPage Calibration Procedures Test Equipment and Setup RequiredFigure A-1. Calibration Setup General Calibration Procedure Are not stored. Exercise care when moving the leads Clear Voltmeter Output Buffer Calibration Program10 ! Calibration Example PauseInput ANY More Outputs to CALIBRATE? Y or N,X$ Disp END of Calibration ProgramFnend Page Page Programming With a Series 200/300 Computer Path NamesVoltage and Current Programming Voltage and Current Programming With Variables Voltage and Current ReadbackService Request and Serial Poll Programming Power Supply RegistersPrint OUTPUT1 is in CV Mode END if Present StatusPrint Overvoltage on Output #1 Enable IntrOFF Intr Print Overvoltage on Output #2Error Detection Programming Outputs Connected In Parallel Stored Operating StatesInput Enter Voltage LIMIT,V Input Enter Operating VOLTAGE,V1Programming Outputs Connected In Series Input Enter the Desired Current Limit POINT,ICommand Summary Table C-1. Command SummaryCommand Description Table C-l. Command Summary PON? ROM?Previously Table D-l. Power-On Self Test Error Message Error Codes and MessagesPower-On Self Test Messages Error Responses Test ResponsesTable D-2. Error Responses Error Code Message Explanation ERR? query ERR keyTable D-3. TEST? Responses Response Code Explanation TEST? queryPage Manual Backdating Make Changes6621A Generally Applicable AnnotationsII. CE92 Product Specific Annotations 6623AFar East Or contactEurope Latin America

6624A, 6621A, 6623A specifications

The HP 6623A, 6621A, and 6624A are precision DC power supplies widely recognized for their reliability and performance in various laboratory and industrial applications. These models are part of HP's 662X series, designed to cater to the needs of engineers, researchers, and technicians who require accurate power sources for their testing and development activities.

The main features of the HP 6623A, 6621A, and 6624A include their high stability, low ripple, and excellent load regulation, ensuring that the output voltage and current remain stable during testing. The power supplies provide multiple output channels, allowing users to power multiple devices simultaneously. The flexibility in setting voltage and current levels makes these models ideal for a wide range of applications, including semiconductor testing, device characterization, and system integration.

One of the standout technologies in the HP 662X series is the use of smart design techniques that minimize noise and enhance output performance. These power supplies incorporate advanced feedback mechanisms to maintain steady output, even under varying load conditions. Additionally, they feature programmable outputs, which means users can adjust the output levels through a connected computer or control system, streamlining the testing process and improving efficiency.

The HP 6623A model offers three independent outputs, with total power capabilities of 40 watts. It includes a 0-20V output, which can deliver up to 2A of current, along with two additional outputs that are adjustable. The 6621A provides a single output option, delivering a maximum of 20V and 2A, making it well-suited for simple applications where a single power source is required. In contrast, the HP 6624A stands out with its four independent output channels, providing a total of 60 watts, making it the most versatile of the three models.

Characteristics of these power supplies include user-friendly interfaces, allowing for easy configuration and monitoring of settings. LED indicators provide real-time feedback on voltage and current levels, enabling users to quickly assess the performance of their tests. Moreover, built-in protection features safeguard both the power supply and the connected devices from overvoltage and overcurrent conditions.

Overall, the HP 6623A, 6621A, and 6624A power supplies embody advanced engineering and design, making them invaluable tools for professionals looking for high-quality, reliable power sources for their electronic testing needs.