HP 6623A Output, Reading the HP-IB Address, Changing the HP-IB Address, Sending a Remote Command

Page 39

OUTPUT

The HP BASIC language statement that addresses the power supply to talk and reads back data from the power supply is:

ENTER

The supply's front panel ADDR annunciator is on when the supply is addressed to talk or to listen.

Reading the HP-IB Address

Before you can operate your power supply remotely, you need to know its HP-IB address. The address was displayed during the power on sequence described in Normal Self Test Indications, page 35. To see the address, press:

ADDR

A typical address display is shown in Figure 3-6:

Figure 3-6. Typical Address Display

The displayed response is the power supply's HP-IB address. When sending a remote command, you append this address to the computer's HP-IB interface select code (normally 7). For example, if the select code is 7 and the power supply's HP- IB address is 5, the combination is 705.

Changing the HP-IB Address

NOTE

All examples in this discussion assume an HP-IB address of 5. It is recommended that you retain this

 

address to simplify programming.

 

 

Every device on the HP-IB must have an address. The supply's address is factory set to decimal 5. Any address from 0 through 30 is a valid address. If you need to change the HP 662xA Supply's address press:

ADDR

You can now enter a new address. For example, press:

1

4

ENTER

You have now changed the address from 5 to 14. If you want to change the address back to 5, repeat the above procedure but use 5 instead of 14 in the last step. Note that the address is stored in the power supply's non-volatile memory and therefore will be retained through interruption of the ac line power.

Sending a Remote Command

To send the power supply a remote command, combine your computer's output statement with the HP-IB interface select code, the HP-IB device address, and finally, the power supply command. For example, to set the output voltage of output channel 1 to 2 volts, send:

Getting Started 39

Image 39
Contents HP Part No Operating ManualCertification Safety Summary Symbol Description Safety SummaryEMC Declaration of ConformityAppendix D--Error Messages Installation ProceduresAppendix A--Calibration Procedure Appendix C--Command SummaryTable Of Contents Local Operation Remote OperationProgramming With The Series 200/300 Computer Command SummaryError Messages CalibrationPage Instrument and Manual Identification Safety ConsiderationsGeneral Information IntroductionAccessories DescriptionOutput Combinations Available Model Basic Operation Output Low Range Values High Range ValuesHP-IB Board Output Boards Specifications Qualifying ConditionsDefinitions Output Response Characteristics Source Effect SpecificationsOutputs Low High Voltage Temperature Coefficient Supplemental CharacteristicsCommand Processing Time see Figure Outputs Low High Voltage Programming ResolutionReadback Resolution AC Input Power and CurrentLow Voltage High Voltage 80 W Low Voltage Safety Agency ComplianceDimensions all models Output ImpedanceGeneral Information General Information General Information General Information Installation Initial InspectionLocation and Cooling Line Fuse Input Power Requirements100/120 2110-0342 220/240 2110-0055 Line FusesLine Voltage Conversion Power CordHP-IB Interface Connector Page Front Panel Controls and Indicators Getting StartedTurning On Your Supply HP-IB Status Annunciators LCL keySystem Control Keys Power Supply StatusAnnunciators Alphanumeric LCDOutput Control Keys Line Switch Numeric Entry KeysNormal Self Test Indications Sample Self-Test Failure Display Checking Out Your Supply Using Local ControlVoltage Test Overvoltage TestCurrent Test Enter/Output Statements Introduction To Remote OperationIset Enter OCPAddr Sending a Remote CommandOutput Reading the HP-IB AddressOften Used Commands Getting Data From The SupplyDisp a Disp a Returning the Supply to Local Mode Output Ranges Output Connections and Operating InformationProtection Features Operating QuadrantsRange Selection Typical Output Range Characteristics Connecting the Load Page Wire Bundled 10 a 20 a FeetCross Section Meters Area in mm2 Remote Voltage Sensing Multiple LoadsPositive and Negative Voltages Remote Sense Connections Remote Voltage SensingOpen Sense Leads Output Type FormulaOutput Noise Considerations Programming Response Time with an Output CapacitorExternal Trigger Circuit Overvoltage Trigger ConnectionsEquivalent Internal OV Trigger Circuit Power Supply Protection Considerations Battery ChargingParallel Operation CV Operation Maximum Allowable Voltage SettingRemote Sensing CC Operation13. Series Connections with Local Sensing CV Operation Series Operation14. Series Connections with Remote Sensing Specifications for Series OperationPage Page Remote Operation HP-IB OperationInterface Function HP-IB Address Selection Power-On Service Request PON Programming SyntaxNumeric Data Sheet 1 of 2. Syntax Forms for Power Supply Commands Sheet 2 of 2. Syntax Forms for Power Supply Commands Ovset Fault ?Vset IsetPage Initial Conditions Power Supply CommandsCurrent Programming Voltage ProgrammingRange Switching Output On/Off Overvoltage OV ProtectionOVSET? Multiple Output Storage & Recall Overcurrent Protection OCPStatus Reporting Clear CommandASTS? UNR +CCBit Assignment of the Serial Poll Register FAULT?UNMASK? Unmask 2,XXXService Request Generation PON ? SRQ?Reprogramming Delay RQS BitDisplay On/Off Other Queries Front Panel ExplanationResponse Code Code Explanation TEST? ResponsesPage General Local ModeLocal Operation Local Control Of Output FunctionsSetting Current Setting VoltageDisplaying the Contents of the Fault Register Setting Overvoltage ProtectionResetting Overvoltage Protection Resetting Overcurrent ProtectionCondition Setting the Reprogramming DelaySetting the Supplys HP-IB Address Local Control Of System FunctionsRCL Enter Displaying Error MessagesAddr Enter STO EnterPage Test Equipment and Setup Required Calibration ProceduresFigure A-1. Calibration Setup General Calibration Procedure Are not stored. Exercise care when moving the leads Pause Calibration Program10 ! Calibration Example Clear Voltmeter Output BufferInput ANY More Outputs to CALIBRATE? Y or N,X$ Disp END of Calibration ProgramFnend Page Page Programming With a Series 200/300 Computer Path NamesVoltage and Current Programming Voltage and Current Readback Voltage and Current Programming With VariablesPresent Status Programming Power Supply RegistersPrint OUTPUT1 is in CV Mode END if Service Request and Serial PollPrint Overvoltage on Output #2 Enable IntrOFF Intr Print Overvoltage on Output #1Error Detection Stored Operating States Programming Outputs Connected In ParallelInput Enter Operating VOLTAGE,V1 Input Enter Voltage LIMIT,VInput Enter the Desired Current Limit POINT,I Programming Outputs Connected In SeriesCommand Summary Table C-1. Command SummaryCommand Description Table C-l. Command Summary ROM? PON?Previously Test Responses Error Codes and MessagesPower-On Self Test Messages Error Responses Table D-l. Power-On Self Test Error MessageError Code Message Explanation ERR? query ERR key Table D-2. Error ResponsesResponse Code Explanation TEST? query Table D-3. TEST? ResponsesPage Make Changes Manual Backdating6623A Generally Applicable AnnotationsII. CE92 Product Specific Annotations 6621ALatin America Or contactEurope Far East

6624A, 6621A, 6623A specifications

The HP 6623A, 6621A, and 6624A are precision DC power supplies widely recognized for their reliability and performance in various laboratory and industrial applications. These models are part of HP's 662X series, designed to cater to the needs of engineers, researchers, and technicians who require accurate power sources for their testing and development activities.

The main features of the HP 6623A, 6621A, and 6624A include their high stability, low ripple, and excellent load regulation, ensuring that the output voltage and current remain stable during testing. The power supplies provide multiple output channels, allowing users to power multiple devices simultaneously. The flexibility in setting voltage and current levels makes these models ideal for a wide range of applications, including semiconductor testing, device characterization, and system integration.

One of the standout technologies in the HP 662X series is the use of smart design techniques that minimize noise and enhance output performance. These power supplies incorporate advanced feedback mechanisms to maintain steady output, even under varying load conditions. Additionally, they feature programmable outputs, which means users can adjust the output levels through a connected computer or control system, streamlining the testing process and improving efficiency.

The HP 6623A model offers three independent outputs, with total power capabilities of 40 watts. It includes a 0-20V output, which can deliver up to 2A of current, along with two additional outputs that are adjustable. The 6621A provides a single output option, delivering a maximum of 20V and 2A, making it well-suited for simple applications where a single power source is required. In contrast, the HP 6624A stands out with its four independent output channels, providing a total of 60 watts, making it the most versatile of the three models.

Characteristics of these power supplies include user-friendly interfaces, allowing for easy configuration and monitoring of settings. LED indicators provide real-time feedback on voltage and current levels, enabling users to quickly assess the performance of their tests. Moreover, built-in protection features safeguard both the power supply and the connected devices from overvoltage and overcurrent conditions.

Overall, the HP 6623A, 6621A, and 6624A power supplies embody advanced engineering and design, making them invaluable tools for professionals looking for high-quality, reliable power sources for their electronic testing needs.