HP 6624A Power Supply Installation Instructions and Guidelines

Page 25

2

Installation

Introduction

This chapter contains instructions for checking and mounting your power supply, connecting your supply to ac power, converting it from one line voltage to another, and connecting the HP-IB cable.

The power supply generates operating magnetic fields which may affect the operation of other instruments. If your instrument is susceptible to magnetic fields, do not locate it in the immediate vicinity of the power supply. Typically, at three inches from the power supply, the electromagnetic field is less than 5 gauss.

NOTE

The HP 662xA power supplies generate operating magnetic fields which may affect the operation of

 

other instruments. If your instrument is susceptible to operating magnetic fields, position it more than 3

 

inches from the HP 662xA supply.

Initial Inspection

Your instrument was thoroughly inspected and tested before it left the factory. As soon as you receive it, remove the power supply from its packing case and check to make sure it has not been damaged in shipment. Check that there are no broken connectors or keys, and that the cabinet and panel surfaces are free from dents and scratches. Check the rear panel terminal blocks and front panel display for any cracks. If damage is found, you should file a claim with the carrier immediately and notify the Hewlett-Packard Sales and Service office nearest you.

Chapter 3 of this manual includes an electrical turn-on check-out procedure which, when carried out successfully, will give you a high level of confidence that the power supply is operating in accordance with its specifications. Detailed electrical checks complete with verification procedures are included in the Service Manual.

Keep the original packing materials for the carrier' s inspection if there was damage, or in case any equipment has to be returned to Hewlett-Packard. Warranty information is printed on the inside cover of this manual. Remember to send a detailed description of the failure and symptoms when returning the power supply for service. Your Hewlett-Packard Sales and Service office will furnish the address of the nearest service office to which the instrument can be shipped .

Location and Cooling

Your power supply can operate without loss of performance within the temperature range of 0 to 55 ° C (measured at the fan intake). The fan, located at the rear of the unit, cools the supply by drawing air in through the openings on the rear panel and exhausting it through openings on the sides. Using Hewlett-Packard rack mount kits will not impede the flow of air.

Because the power supply is fan cooled, it must be installed in a location that allows sufficient space at the rear and the sides for adequate circulation of air. Either side may be restricted to have as little as 1 inch (25 mm) space.

Figure 2-1 gives the dimensions of the power supply cabinet. These dimensions apply to all five models. The cabinet has plastic feet that are shaped to ensure self-alignment when stacked with other Hewlett-Packard System II cabinets. The feet may be removed for rack mounting.

The power supply can be mounted in a standard 19 inch rack panel or enclosure. Rack mounting accessories for this unit are listed on page 12, under Options of Chapter 1. Complete installation instructions are included with each rack mounting kit. Instrument support rails are required for non-stationary installations. These are normally supplied with the cabinet and are not included with the rack mounting kits.

Installation 25

Image 25
Contents HP Part No Operating ManualCertification Safety Summary Symbol Description Safety SummaryEMC Declaration of ConformityAppendix A--Calibration Procedure Installation ProceduresAppendix C--Command Summary Appendix D--Error MessagesTable Of Contents Local Operation Remote OperationError Messages Command SummaryCalibration Programming With The Series 200/300 ComputerPage General Information Safety ConsiderationsIntroduction Instrument and Manual IdentificationDescription AccessoriesOutput Combinations Available Model Output Low Range Values High Range Values Basic OperationHP-IB Board Output Boards Qualifying Conditions SpecificationsDefinitions Output Response Characteristics Source Effect SpecificationsOutputs Low High Voltage Temperature Coefficient Supplemental CharacteristicsReadback Resolution Outputs Low High Voltage Programming ResolutionAC Input Power and Current Command Processing Time see FigureDimensions all models Safety Agency ComplianceOutput Impedance Low Voltage High Voltage 80 W Low VoltageGeneral Information General Information General Information General Information Initial Inspection InstallationLocation and Cooling Line Fuse Input Power Requirements100/120 2110-0342 220/240 2110-0055 Line FusesLine Voltage Conversion Power CordHP-IB Interface Connector Page Getting Started Front Panel Controls and IndicatorsTurning On Your Supply HP-IB Status Annunciators LCL keyAnnunciators Power Supply StatusAlphanumeric LCD System Control KeysOutput Control Keys Numeric Entry Keys Line SwitchNormal Self Test Indications Sample Self-Test Failure Display Checking Out Your Supply Using Local ControlOvervoltage Test Voltage TestCurrent Test Iset Enter Introduction To Remote OperationOCP Enter/Output StatementsOutput Sending a Remote CommandReading the HP-IB Address AddrGetting Data From The Supply Often Used CommandsDisp a Disp a Returning the Supply to Local Mode Output Ranges Output Connections and Operating InformationOperating Quadrants Protection FeaturesRange Selection Typical Output Range Characteristics Connecting the Load Page Feet Wire Bundled 10 a 20 aCross Section Meters Area in mm2 Multiple Loads Remote Voltage SensingPositive and Negative Voltages Remote Sense Connections Remote Voltage SensingOutput Noise Considerations Output Type FormulaProgramming Response Time with an Output Capacitor Open Sense LeadsExternal Trigger Circuit Overvoltage Trigger ConnectionsEquivalent Internal OV Trigger Circuit Battery Charging Power Supply Protection ConsiderationsParallel Operation CV Operation Maximum Allowable Voltage SettingRemote Sensing CC Operation13. Series Connections with Local Sensing CV Operation Series Operation14. Series Connections with Remote Sensing Specifications for Series OperationPage Page HP-IB Operation Remote OperationInterface Function HP-IB Address Selection Programming Syntax Power-On Service Request PONNumeric Data Sheet 1 of 2. Syntax Forms for Power Supply Commands Sheet 2 of 2. Syntax Forms for Power Supply Commands Vset Fault ?Iset OvsetPage Initial Conditions Power Supply CommandsCurrent Programming Voltage ProgrammingRange Switching Overvoltage OV Protection Output On/OffOVSET? Multiple Output Storage & Recall Overcurrent Protection OCPStatus Reporting Clear CommandASTS? UNR +CCUNMASK? FAULT?Unmask 2,XXX Bit Assignment of the Serial Poll RegisterService Request Generation PON ? SRQ?RQS Bit Reprogramming DelayDisplay On/Off Other Queries Explanation Front PanelResponse Code Code Explanation TEST? ResponsesPage Local Operation Local ModeLocal Control Of Output Functions GeneralSetting Current Setting VoltageResetting Overvoltage Protection Setting Overvoltage ProtectionResetting Overcurrent Protection Displaying the Contents of the Fault RegisterSetting the Supplys HP-IB Address Setting the Reprogramming DelayLocal Control Of System Functions ConditionAddr Enter Displaying Error MessagesSTO Enter RCL EnterPage Test Equipment and Setup Required Calibration ProceduresFigure A-1. Calibration Setup General Calibration Procedure Are not stored. Exercise care when moving the leads 10 ! Calibration Example Calibration ProgramClear Voltmeter Output Buffer PauseDisp END of Calibration Program Input ANY More Outputs to CALIBRATE? Y or N,X$Fnend Page Page Path Names Programming With a Series 200/300 ComputerVoltage and Current Programming Voltage and Current Readback Voltage and Current Programming With VariablesPrint OUTPUT1 is in CV Mode END if Programming Power Supply RegistersService Request and Serial Poll Present StatusOFF Intr Enable IntrPrint Overvoltage on Output #1 Print Overvoltage on Output #2Error Detection Stored Operating States Programming Outputs Connected In ParallelInput Enter Operating VOLTAGE,V1 Input Enter Voltage LIMIT,VInput Enter the Desired Current Limit POINT,I Programming Outputs Connected In SeriesTable C-1. Command Summary Command SummaryCommand Description Table C-l. Command Summary ROM? PON?Previously Power-On Self Test Messages Error Responses Error Codes and MessagesTable D-l. Power-On Self Test Error Message Test ResponsesError Code Message Explanation ERR? query ERR key Table D-2. Error ResponsesResponse Code Explanation TEST? query Table D-3. TEST? ResponsesPage Make Changes Manual BackdatingII. CE92 Product Specific Annotations Generally Applicable Annotations6621A 6623AEurope Or contactFar East Latin America

6624A, 6621A, 6623A specifications

The HP 6623A, 6621A, and 6624A are precision DC power supplies widely recognized for their reliability and performance in various laboratory and industrial applications. These models are part of HP's 662X series, designed to cater to the needs of engineers, researchers, and technicians who require accurate power sources for their testing and development activities.

The main features of the HP 6623A, 6621A, and 6624A include their high stability, low ripple, and excellent load regulation, ensuring that the output voltage and current remain stable during testing. The power supplies provide multiple output channels, allowing users to power multiple devices simultaneously. The flexibility in setting voltage and current levels makes these models ideal for a wide range of applications, including semiconductor testing, device characterization, and system integration.

One of the standout technologies in the HP 662X series is the use of smart design techniques that minimize noise and enhance output performance. These power supplies incorporate advanced feedback mechanisms to maintain steady output, even under varying load conditions. Additionally, they feature programmable outputs, which means users can adjust the output levels through a connected computer or control system, streamlining the testing process and improving efficiency.

The HP 6623A model offers three independent outputs, with total power capabilities of 40 watts. It includes a 0-20V output, which can deliver up to 2A of current, along with two additional outputs that are adjustable. The 6621A provides a single output option, delivering a maximum of 20V and 2A, making it well-suited for simple applications where a single power source is required. In contrast, the HP 6624A stands out with its four independent output channels, providing a total of 60 watts, making it the most versatile of the three models.

Characteristics of these power supplies include user-friendly interfaces, allowing for easy configuration and monitoring of settings. LED indicators provide real-time feedback on voltage and current levels, enabling users to quickly assess the performance of their tests. Moreover, built-in protection features safeguard both the power supply and the connected devices from overvoltage and overcurrent conditions.

Overall, the HP 6623A, 6621A, and 6624A power supplies embody advanced engineering and design, making them invaluable tools for professionals looking for high-quality, reliable power sources for their electronic testing needs.