Omega Vehicle Security FPU500 manual Variances in Normalized Flow Rate mL per Revolution

Page 43

Specifications 9

 

Table 9-2. Average Flow Rates (Cont’d)

 

 

 

 

 

 

Tubing Wall

Tubing Size

mL per

Minimum Flow

Maximum Flow

Thickness

Inner Diameter

Revolution

Rate at 600 RPM

Rate at 600 RPM

 

 

 

(mL/Minute)

(mL/Minute)

1.5 mm

3.0 mm

0.8*

8

482

1.5 mm

4.0 mm

1.43*

15

857

1.5 mm

5.0 mm

2.1*

21

1257

1.5 mm

6.0 mm

2.7*

27

1607

1.5 mm

7.0 mm

3.6*

37

2187

1.5 mm

8.0 mm

3.9*

39

2316

* Metric mL per Revolution numbers are rounded off

Normalized flow rates (mL per revolution) vary significantly, based on motor speed, tubing materials, and mechanical tolerances in pump dimensions. In addition, the variances are different for the different tubing dimensions. Variances from the normalized flow rates of Table 9-2 are shown below. The data is shown for new tubing only.

Tubing operating life tests are done at 600 RPM, with 20°C water, 0 PSI back pressure until the tubing breaks

Table 9-3. Variances in Normalized Flow Rate (mL per Revolution)

Tubing Size

Due to

Between Pumps

Due to Different

Inner Diameter

Motor

(due to Mechanical

Tubing Materials

(in.)

Speed

Tolerances)

 

 

 

 

 

132

±10%

±25%

±20%

116

±10%

±15%

±15%

18

±10%

±10%

±5%

316

±5%

±10%

±5%

14

±5%

±15%

±5%

516

±15%

±20%

±10%

9-3

Image 43
Contents User’s Guide Servicing North America From the Technical Library Peristaltic Pump Index Index Introduction FeaturesDescription Introduction Overall View of the Pump Parts of the PumpDescription Function View of the Left Side Left Side of the PumpView of the Right Side Right Side of the PumpParts of the Pump Introduction Setting Up the PumpsRequired Hardware Attaching a Single Pump Directly to the Motor Attaching the PumpParts of the Pump Stacking Two Pumps onto One Motor Figure B Attaching a Single Pump to an Adapter Plate Mounting the adapter plate to the motor refer to Figure Using Your Own Adapter Plate Setting Up the Pumps Selecting Tubing Tubing InformationTubing Life Tubing Information Operating the Pump Opening the Stator Loading the TubingFigure D 5a. Inlet and Outlet Flow Operating the Pump5b. Inlet and Outlet Flow Adjusting the Clamp ScrewTurn the pump motor off Operating the Pump Maintenance Shows the exploded view of the pump Replacing the Rotor Assembly Part Number FPU500-RAMaintenance Maintenance Problem Solution Troubleshooting GuideProblem Check motor fuse Troubleshooting Guide Design Considerations Design SolutionTechnical Details Theory of OperationStacking More than Two Pumps Specifications ⁄16 ⁄32 132 540 1140 1800 2280 210 Average Flow Rates Cont’d Variances in Normalized Flow Rate mL per RevolutionTubing Average Average Tubing LifeAverage Flow Rates for Viscous Liquids Average Tubing Life ContdSpecifications Spare Parts Spare Parts and AccessoriesPart Number Description Viton Peristatic Pump Motor 10-4 WARRANTY/DISCLAIMER Shop online at omega.com