Milwaukee 2216-20 manual Calibration, Repair Service, Instructions, Environmental, Condition

Page 1
Environmental

MILWAUKEE ELECTRIC TOOL CORPORATION – TEST & MEASUREMENT PRODUCT

REPAIR SERVICE INSTRUCTIONS — CALIBRATION 2216-20 Light Commercial Digital Multimeter

 

Perform all calibration at an ambient temperature of 23°C ± 2°C / 73.4°F ± 3.6°F and relative

Environmental

humidity of 80% - allow the digital multimeter to sit at this temperature for a minimum of 30

Condition

minutes before proceeding.

 

REPAIR SERVICE

INSTRUCTIONS

Calibration

MILWAUKEE ELECTRIC TOOL CORPORATION

13135 WEST LISBON ROAD ● BROOKFIELD, WISCONSIN 53005-2550 ● USA

58-92-2216d1 10/2010

Image 1
Contents Calibration REPAIR SERVICEINSTRUCTIONS MILWAUKEE ELECTRIC TOOL CORPORATIONPage TABLE OF CONTENTSThe Symbols Precautions and Safety InformationIntroduction Specific precautions SAFETYGeneral specifications SPECIFICATIONSCharacteristics DescriptionMeasurement Characteristics 3 DC Current 6 Continuity 15Low Battery Voltage 2.3V ± 16Backlight 11Auto Power Off17Battery life ALKALINE 12Measuring Rate 3 times/secCharacteristics Physical and Environmental CharacteristicsEnvironmental characteristics DescriptionRequired Equipment Replacing the Battery Basic MaintenanceOpening the Meter Case Testing Fuses WarningFuse Replacement CleaningPerformance Tests Testing the DisplayTesting the Voltage Function Testing the Resistance Function Testing the Ampere Function Step Table 7 Frequency TestSource LevelEnvironmental Condition Calibration ProcedureCalibration Procedure AOpen the bottom caseEDCA Calibration VR7 2.Tighten the screwFResistance 6kΩ Range Calibration VR6 GCapacitance Calibration VR1