Milwaukee 2216-20 manual Calibration Procedure, Environmental Condition, AOpen the bottom case

Page 19

Calibration Procedure

Recalibrate the meter:

It is recommended that the meter may be calibrated once year. Use the following procedure to calibrate the meter.

Environmental Condition

1.Perform calibration at an ambient temperature and relative humidity (23°C±2°C and R.H.<=80%). Allow the instrument to sit at this temperature for at least thirty minutes.

Calibration Procedure

(A)Open the bottom case

1.Disconnect the test leads from any circuit under test and turn off the meter.

2.Loosen the screw from the bottom case.

3.Remove the bottom case.

4.Supply 3V to the meter.

(B)DCV Calibration (VR2)

1.Rotate the rotary switch to DCV Function.

2.Connect the output of a DCV calibrator to V and COM input terminals of the meter.

3.Output DCV 4.000V from the calibrator.

4.Adjust VR2 until the display shows 4.000V ± 0.001V.

5.Turn off the output of the calibrator and disconnect the meter from the calibrator.

(C)ACV Calibration (VR3)

1.Rotate the rotary switch to ACV Function.

2.Connect the output of an ACV calibrator to V and COM input terminals of the meter.

3.Output ACV 4.000V @ 60Hz from the calibrator.

4.Adjust VR3 until the display shows 4.015V ± 0.001V.

5.Turn off the output of the calibrator and disconnect the meter from the calibrator.

(D)DCV mV Calibration (VR4)

1.Rotate the rotary switch to DCV Function.

2.Connect the output of a DCV calibrator to V and COM input terminals of the meter.

3.Output DCV 400.0mV from the calibrator.

4.Adjust VR4 until the display shows 400.0mV ± 0.1mV.

5.Turn off the output of the calibrator and disconnect the meter from the calibrator.

17

Image 19
Contents MILWAUKEE ELECTRIC TOOL CORPORATION REPAIR SERVICECalibration INSTRUCTIONSPage TABLE OF CONTENTSThe Symbols Precautions and Safety InformationIntroduction Specific precautions SAFETYDescription SPECIFICATIONSGeneral specifications CharacteristicsMeasurement Characteristics 3 DC Current 6 Continuity 12Measuring Rate 3 times/sec 11Auto Power Off15Low Battery Voltage 2.3V ± 16Backlight 17Battery life ALKALINEDescription Physical and Environmental CharacteristicsCharacteristics Environmental characteristicsRequired Equipment Replacing the Battery Basic MaintenanceOpening the Meter Case Cleaning WarningTesting Fuses Fuse ReplacementPerformance Tests Testing the DisplayTesting the Voltage Function Testing the Resistance Function Testing the Ampere Function Level Table 7 Frequency TestStep SourceAOpen the bottom case Calibration ProcedureEnvironmental Condition Calibration ProcedureGCapacitance Calibration VR1 2.Tighten the screwEDCA Calibration VR7 FResistance 6kΩ Range Calibration VR6