Milwaukee 2216-20 manual Testing the Resistance Function

Page 16
Testing the Resistance Function

Testing the Resistance Function

To verify the accuracy of the resistance function, do the following:

1.Connect the Calibrator to VΩHz and COM on the Meter.

2.Turn the rotary switch to Ω.

3.Apply the inputs for step 1-6 in Table 3.

4.Compare the Meter display readings to the display readings in Table 3.

5.If the display reading falls outside of the range shown in Table 3, the Meter does not meet specification.

Table 3 Resistance Test :

Step

Source

Reading

1

560.0

553.9 to 566.1

 

 

 

2

5.600k

5.539k to 5.661k

 

 

 

3

56.00k

55.39k to 56.61k

 

 

 

4

560.0k

553.9k to 566.1k

 

 

 

5

5.600M

5.539M to 5.661M

 

 

 

6

36.00M

35.44M to 36.56M

 

 

 

*Lead resistance on the 400Ω range is not included in error.

Testing the Capacitance Function

The Meter measures capacitance by charging the capacitor with a known Direct current, measuring the resultant voltage, and calculating the capacitance. If the same capacitance is measured on an impedance bridge, a different reading may result. This variance is likely to be Greater at higher frequencies.

To verify the accuracy of the capacitance measuring function, do the Following:

1.Apply the Capacitor to the VΩHz and COM inputs on the Meter. For steps 1-4 in Table 4.

2.Turn the rotary switch to Manual backgroundManual background

3.Compare the reading on the Meter display to the reading in Table 4.

4.If the display reading falls outside of the range shown in Table 4, the Meter does not meet specification.

Table 4 Capacitance Test :

Step

Source

Reading

1

3.6uF

3.3uF to 3.9uF

 

 

 

2

80.0uF

78.3uF to 81.7uF

 

 

 

3

320uF

312uF to 328uF

 

 

 

4

1000uF

979uF to 1021uF

 

 

 

14

Image 16
Contents REPAIR SERVICE CalibrationINSTRUCTIONS MILWAUKEE ELECTRIC TOOL CORPORATIONTABLE OF CONTENTS PageThe Symbols Precautions and Safety InformationIntroduction SAFETY Specific precautionsSPECIFICATIONS General specificationsCharacteristics DescriptionMeasurement Characteristics 3 DC Current 6 Continuity 11Auto Power Off 15Low Battery Voltage 2.3V ± 16Backlight17Battery life ALKALINE 12Measuring Rate 3 times/secPhysical and Environmental Characteristics CharacteristicsEnvironmental characteristics DescriptionRequired Equipment Replacing the Battery Basic MaintenanceOpening the Meter Case Warning Testing FusesFuse Replacement CleaningTesting the Display Performance TestsTesting the Voltage Function Testing the Resistance Function Testing the Ampere Function Table 7 Frequency Test StepSource LevelCalibration Procedure Environmental ConditionCalibration Procedure AOpen the bottom case2.Tighten the screw EDCA Calibration VR7FResistance 6kΩ Range Calibration VR6 GCapacitance Calibration VR1